Check the message system of the amalgam separator

An electronic control device monitors the function of the amalgam
separator. It detects any mechanical blocking or failure of the drive
motor. The error is indicated by an acoustic signal.
The function of this message system must be checked at least once per
year.
1. Start the amalgam test in the setup dialog "Maintenance/Service
functions/Water unit  [→ 190]/Error signal of the amalgam separator"
by long pressing the "Start test" key.
Ä A buzzer tone should sound. The amalgam test remains active
for 10 seconds.
2. During this time, open the Notifications setup dialog.
Ä As long as the amalgam test is active, the message "Amalgam
separator 100 % full - Replace amalgam separator" should
appear.
Ä The display and message systems are OK when a buzzer tone
sounds and the message appears.
Inform your service technician if the error occurs during regular
operation.
In Germany: Document the test in the operation log of the amalgam
separator in the "Testing the function of display and message systems"
section.