NOTICE
Protecting the motor shaft/axis
A bent motor shaft on the BL E motor causes irregular operating
noises or strong vibrations. This can damage instruments.
➢ Do not let the motor fall to the floor.
NOTICE
Never lubricate electric motors!
➢ Remove the handpieces from the electric motors at the end of the
working day so that no oil can run into the motor overnight.
4.5.11.5

Connecting the instrument hose

ü Colored marks and position of guide lugs on motor and hose
coupling of the instrument hose match.
1. Slide back the cap nut at the hose coupling.
2. Attach the motor onto the hose coupling up to the stop, observing
the contact pins and tubes. Make sure the hose coupling does not
tilt.
Ä The arrow on the hose coupling and the notch on the motor
must face each other.
3. Press the cap nut gently onto the thread; then turn it
counterclockwise until a faint click is heard.
4. Screw the cap nut tightly onto the motor clockwise.
Does water leak out between the motor and the hose connection?
1. Remove the motor from the instrument hose.
2. Reconnect the motor to the instrument hose. Make sure it is
connected properly.
3. If water still leaks, replace the sealing washer.
4.5.11.6

Changing an instrument

WARNING
The instrument should be attached or removed only when the motor
is at standstill.
Attach the instrument/adapter to the BL E motor
WARNING
Do not operate the BL E motor with an exposed motor shaft and
attachments (removed instrument/adapter). Risk of injury!
ü The motor is at a standstill.
➢ Attach the instrument or adapter. Lock the instrument or adapter
into place by turning it.
4 Operation
4.5 Dentist element
101