Patient chair

4.4.1

Safety instructions

WARNING
The free space under the patient couch and up to the water unit can
be decreased due to chair movements.
Parts of the patient's or user's body may be pinched or crushed.
➢ Do not allow any limbs to stick out in the space between the chair
upholstery, armrests and chair base. Please make sure that the
patient's arms and legs rest on the upholstery of the chair.
➢ Do not place any objects on the base of the chair.
CAUTION
The maximum load capacity of the patient chair is 185 kg (407.9 lbs)
according to ISO 7494-1 (tested with multiple safety according to IEC
60601-1).
If the maximum load capacity is exceeded, there is a risk of damage
to the treatment chair and injury to the patient.
➢ Never allow any persons who weigh more than 180 kg (396.8 lbs)
to sit on the patient chair. The maximum permitted load capacity
is indicated on a label next to the rating plate of the treatment
center.
➢ The maximum additional weight of accessories mounted on the
patient chair is 5 kg (11 lbs).
WARNING
Objects may protrude into the movement range of the chair.
There is a risk of crushing the patient and damaging the objects.
➢ Make sure that no objects such as e.g. windows, drawers or other
devices protrude into the movement range of the treatment chair.
IMPORTANT
Chair interlock
As long as a treatment instrument is activated, all functions for
moving the patient chair are disabled for safety reasons.
If chair movement is permanently blocked, please contact your service
technician.