5 Care, cleaning and maintenance by the practice team
When the treatment center is first switched on during the working day,
this is interpreted to be autopurge mornings. When it is switched off the
same day, this is detected as autopurge evenings.
Insert water-carrying instruments and suction hoses into the water
unit
The water unit has integrated sanitization adapters for water-carrying
instruments and adapters for suction hoses. They enable all instruments
to be purged with water. For this purpose, water-carrying instruments
must be inserted into the sanitization adapters and the suction hoses
into the adapters for suction hose cleaning.
ü The Remove instruments message is displayed.
1. Remove the Sprayvit sleeves from the valve bodies, the straight
and contra-angle handpieces from the water-carrying instruments,
and the suction tips from the suction hoses.
2. Dentist element side: Plug the couplings of all water-carrying
treatment instruments into the sanitization adapters in the water unit
(for Sprayvit: valve lever pointing up, locking knob pointing down).
Always ensure that the instruments are positioned correctly to avoid
errors during the flow rate check or leakages.
IMPORTANT
Colors of the sanitization adapters
The sanitization adapters for the instrument couplings are color
coded. The assignment is also displayed on the Smart Touch.
Yellow on left = Sprayvit of the assistant element
Yellow on right = Sprayvit of the dentist element
White = turbine
Green = BL E motor
Blue = BL ISO F motor (ISO interface)
Red = SiroSonic TL ultrasonic handpiece
IMPORTANT
Water button of the Sprayvit
Depending on the position of the Sprayvit water button, the removable
sanitization cartridge is marked on the back. When the nozzle is taken
from the holder, this mark can also be seen on the Sprayvit. If the
water button is on the right, the marking is an "R", for the Sprayvit
with water button on the left "L". The marking on the removable
sanitization cartridge must match that on the Sprayvit.
3. Assistant element side: Attach the suction hoses to the adapters.
When using two large and one small suction hose, the small suction
hose must be placed in the middle.
Ä When an instrument has been taken from the holder, this is
displayed on the Smart Touch by a highlighted instrument and a
gray circle around the instrument position.
5.3 Instruments and instrument hoses