4 Operation
4.4 Patient chair
70
4.4.7.1

Moving the patient chair to the entry/exit position

The following functions are triggered to facilitate patient entry and exit in
the entry/exit position:
● The patient chair moves to an upright position
● The operating light switches off
● The cuspidor swivels out (optional)
For the chair program entry/exit position (0), the following can be
configured:
● The position of the patient chair can be changed, see "Configuring
chair programs and shock positioning"  [→ 76].
● The tumbler heater can be switched off automatically, see the
"Settings/Hygiene/Tumbler  [→ 176]/Water heater" setup dialog.
● The image transmitted to the patient monitor can be automatically
replaced by a default display, see "Settings/General/Patient monitor 
[→ 166]" setup dialog.
CAUTION
The patient's feet may get caught in the instrument hoses of the
dentist element when they enter or leave the patient chair.
The patient may trip or fall.
➢ Turn the dentist element outward before the patient enters or
leaves it.
Via Smart Touch
ü The Start dialog is displayed on the Smart Touch.
➢ Tap on the 0 key (< 2 s).
Via foot control
ü All instruments are in their holders.
➢ Press the right button of the foot control.
Via the assistant element
➢ Press the 0 key on the assistant element briefly (< 2 s).
Via the 4-way foot switch
In the setup dialog, you can set whether the chair program enter/exit
position (0) can be moved to via the 4-way foot switch, see "Settings/
Operation/4-way foot switch  [→ 172]". The spray aspirator must be put
in the holder if it has been set in the setup dialog that the suction flow
can be switched on/off via the 4-way foot switch.
Instructions for Use Treatment Center Axano Pure
D3685.201.02.01.02    2023-12
Dentsply Sirona
68 29 514 D3685