Dentsply Sirona Treatment Center Axano Pure Instructions For Use Manual page 260

Table of Contents

Advertisement

5 Care, cleaning and maintenance by the practice team
5.6 Sanitization
260
➢ Check the instruments in the instrument positions marked with a red
exclamation mark.
Ä When all of the instruments have been put in the holders, the
sanitization program will continue.
Insert water-carrying instruments and suction hoses into the water
unit
The water unit has integrated sanitization adapters for water-carrying
instruments and adapters for suction hoses. This allows all instruments
and suction hoses to be simultaneously treated with a high
concentration of sanitization solution and then rinsed with water. For
this purpose, sanitizable instruments must be inserted in the sanitization
adapters and suction hoses must be inserted into the adapters for
suction hose cleaning.
ü The message "Put all rinsable instruments into the colored adapters
and press start" is displayed.
1. Detach the Sprayvit sleeves from the valve bodies, the straight and
contra-angle handpieces from the instruments that can be sanitized,
and the suction tips from the suction hoses.
2. Dentist element side: Insert the couplings of all water-carrying
treatment instruments into the sanitization adapters in the water unit
(for Sprayvit: valve lever pointing up, locking knob pointing down).
Always ensure that the instruments are seated correctly to avoid
errors during the flow rate check or leakages.
IMPORTANT
Colors of the sanitization adapters
The sanitization adapters for the instrument couplings are color
coded. The assignment is also displayed on the Smart Touch.
Yellow on left = Sprayvit of the assistant element
Yellow on right = Sprayvit of the dentist element
White = turbine
Green = BL E motor
Blue = BL ISO F motor (ISO interface)
Red = SiroSonic TL ultrasonic handpiece
IMPORTANT
Water button of the Sprayvit
Depending on the position of the Sprayvit water button, the removable
sanitization cartridge is marked on the back. When the nozzle is taken
from the holder, this mark can also be seen on the Sprayvit. If the
water button is on the right, the marking is an "R", for the Sprayvit
with water button on the left "L". The marking on the removable
sanitization cartridge must match that on the Sprayvit.
Instructions for Use Treatment Center Axano Pure
D3685.201.02.01.02    2023-12