4 Operation
4.5 Dentist element
124
3. Tap on the Cal key.
Ä The button flashes orange. During calibration, increasing
speeds are set on the motor. When the bur drive is calibrated,
the key is permanently marked orange. Calibration is then
finished.
4.5.14.2.3
Select favorite implantology system
ü The implantology dialog is displayed on the Smart Touch.
1. Tap on the Select treatment key.
Ä A list of favorite implantology systems is displayed.
2. Select one of the displayed implantology systems. In the submenu,
various diameters can be selected. If the desired system is not on
the list, it has not yet been marked as a favorite, see "Managing
implantology systems"  [→ 128].
Ä After tapping on an implantology system, the list is hidden. The
implantology dialog displays the instruments of the selected
system.
4.5.14.2.4
Selecting implant instrument
ü The instruments of the selected implantology system are displayed
in the implantology dialog.
➢ Select the desired implant instrument. Tap on the Instrument
selection keys or one of the instruments displayed. Swipe left and
right to display other instruments for the selected implantology
system.
or
➢ The instrument selection is the starting position for the red cursor.
Press the left or right swivel button of the foot control.
Ä The selected instrument is marked.
Ä The speed and torque values specified by the manufacturer are
displayed.
WARNING
The parameters of the implantology instruments can be changed by
the respective manufacturer. Therefore, please make sure to always
check the specifications of the respective manufacturer prior to use.
You can return to the start of implantology treatment using the Restart
treatment key. The first implant instrument is selected.
4.5.14.2.5
Changing speed and torque
If you do not want to use the stored standard parameters of the implant
instruments, you can change them.
In implantology treatment, the speed and torque values of the contra-
angle handpiece, and not those of the motor, are specified. The control
electronics of the bur drive calculate motor control based on the
specified gear reduction and the speed and torque settings.