4.5.14.1.10.3
Apex and endodontics beeps
Apex beeps
In addition to the graphical distance display on the Smart Touch, the
position of the file in the root canal can also be indicated as an acoustic
signal.
● No beep is played if the file is at least five display levels from the
apex on the distance display.
● Beeps with long intervals are played if the file is three or four display
levels away from the apex.
● Beeps with short intervals are played if the file is one or two display
levels away from the apex.
● A continuous beep is played when the file has reached or passed
the apex.
The volume of the beeps for apex measurement can be set in three
levels, see "Settings/Operation/Sounds  [→ 174]/Apex locator". You can
also switch off the beeps.
Tip: In the expanded endodontics dialog, keys can also be placed to
adjust the volume of the Apex beeps, see "User interface  [→ 183]".
Endodontics beeps
The following endodontics beeps are emitted:
● One beep sounds when the set torque value is reached.
● A beep sounds if the physiological apex is reached when the motor
stop function is switched off.
● A beep sounds when the motor stops automatically once it has
reached the preset motor stop position (loudspeaker symbol on the
distance display).
● Three beeps sound when the motor switches to counterclockwise
rotation if the auto-reverse function is activated and the foot pedal is
pressed again.
The endodontics signals can be switched on/off in the Setup dialog, see
"Settings/Operation/Sounds  [→ 174]/Endodontics sounds".