Lubricate, disinfect/sterilize the treatment instruments

5.3.3.1

Treatment instruments with separate instructions for use

WARNING
The procedures required for the following treatment instruments are
described in the following separate instructions for use:
• Straight and contra-angle handpieces in various versions
• Turbines
• SiroSonic TL ultrasonic handpiece
5.3.3.2
Cleaning, disinfecting/sterilizing the Sprayvit multifunctional
syringe
NOTICE
Approved care, cleaning, and disinfecting agents
Use only care, cleaning, and disinfecting agents that have been
approved by Dentsply Sirona, see "Care, cleaning, and disinfecting
agents"  [→ 198]!
All disinfectants must be approved in your country and have proven
bactericidal, fungicidal and virucidal properties. Use only disinfectants
with no protein-fixing properties.
Do not use agents that are strongly acidic ,(pH < 5), alkaline (pH > 9)
or contain a lot of chlorine.
NOTICE
Never clean in an ultrasonic bath!
Never immerse in disinfectants!
After each treatment session
NOTICE
Perform reprocessing immediately after treatment, after 1 hour at the
latest.
ü Wear suitable protective clothing.
1. Remove any residue, e.g., from impression material or caustic
chemicals, immediately.
2. Purge the water and air channels on the treatment center for 30
seconds.
3. Predisinfect directly at the treatment center  [→ 220].
4. Remove the nozzle and the housing  [→ 93].
5. Transport the nozzle and housing to the hygiene room in a suitable
transport container.
6. Perform automated reprocessing  [→ 220]. Manual reprocessing 
[→ 221] is possible in exceptional cases if national/local regulations
are followed.
7. If necessary, sterilize the housing, keyboard and nozzle  [→ 221].