Treatment Center
Axano Pure
Instructions for Use
English

Advertisement

Table of Contents
loading

Summary of Contents for Dentsply Sirona Treatment Center Axano Pure

  • Page 1 2023-12 New as of: Treatment Center Axano Pure Instructions for Use English...
  • Page 2: Table Of Contents

    Table of contents Dentsply Sirona Instructions for Use Treatment Center Axano Pure Table of contents General information..................Dear customer,......................Contact information ..................... General information about these instructions for use..........Scope of these instructions for use ................Other valid documents ....................
  • Page 3 Dentsply Sirona Table of contents Instructions for Use Treatment Center Axano Pure Unit description....................Operating principle ...................... Standards/Approvals ....................Technical data ......................System overview ......................Patient chair ........................ Motor-driven headrest ....................Double-jointed head support ..................Foot control ......................... Smart Control....................
  • Page 4 Table of contents Dentsply Sirona Instructions for Use Treatment Center Axano Pure 4.3.3 Using the cursor control ................4.3.3.1 Functionality................. 4.3.3.2 Using cursor control with Comfort foot control......4.3.3.3 Using cursor control with Classic foot control ......Patient chair ........................
  • Page 5 Dentsply Sirona Table of contents Instructions for Use Treatment Center Axano Pure 4.5.8 General instrument functions ................ 4.5.8.1 Selecting a coolant ..............4.5.8.2 Switching the preselected coolant on/off ........4.5.8.3 Activating the chip blower ............4.5.8.4 Setting the amount of spray water ..........
  • Page 6 Table of contents Dentsply Sirona Instructions for Use Treatment Center Axano Pure 4.6.3.4 Composite function ..............132 4.6.3.5 Chair programs ................133 4.6.3.6 Headrest ..................133 4.6.3.7 Timer.................... 133 4.6.3.8 Bell....................133 4.6.3.9 Function key ................133 4.6.4 Suction handpieces..................134 SmartLite Pro curing light................
  • Page 7 Dentsply Sirona Table of contents Instructions for Use Treatment Center Axano Pure 4.16.5 Settings ......................164 4.16.5.1 General ..................165 4.16.5.2 Treatment chair................167 4.16.5.3 Instruments .................. 167 4.16.5.4 Operation ..................170 4.16.5.5 Operating light ................175 4.16.5.6 Hygiene..................176 User settings................
  • Page 8 Table of contents Dentsply Sirona Instructions for Use Treatment Center Axano Pure 5.3.3.3 Disinfecting/sterilizing motors and adapters ........ 222 5.3.3.4 Clean and disinfect/sterilize the components of the ApexLocator 227 5.3.3.5 Disinfecting/sterilizing the SmartLite Pro curing light....227 5.3.3.6 Cleaning/disinfecting the SiroCam AF / AF+ intraoral camera ..228 5.3.4...
  • Page 9 Dentsply Sirona Table of contents Instructions for Use Treatment Center Axano Pure Disposal......................273 Battery disposal......................274 Disposal of disinfectants....................274 Overview of all function keys................275 Fixed Keys........................275 9.1.1 Dentist element ..................... 275 9.1.2 Assistant element..................276 Start dialog ........................277 Instrument dialog......................
  • Page 10: General Information

    1 General information Dentsply Sirona 1.1 Dear customer, Instructions for Use Treatment Center Axano Pure General information Dear customer, We are pleased that you have equipped your practice with the Axano Lite treatment center from Dentsply Sirona. Our aim is to identify our customers' needs early and create innovative solutions.
  • Page 11: General Information About These Instructions For Use

    The instructions as well as all warning and safety notices they contain must also be followed. The following Dentsply Sirona products have separate instructions for use: ● Straight and contra-angle handpieces ● Turbines ●...
  • Page 12: Warranty And Liability

    Exclusion of liability If the operator does not meet the obligation to carry out such maintenance or error messages are ignored, Dentsply Sirona or its authorized dealer does not assume any liability for resulting damage. Usage data The treatment center saves usage data for quality and product improvement and for efficient remote support when servicing is needed.
  • Page 13: Indications For Use

    Dentsply Sirona 1 General information Instructions for Use Treatment Center Axano Pure 1.9 Indications for use Indications for use The Treatment Center is suitable for indications from the following specialist fields of human dentistry: ● Preventive dentistry ● Restorative dentistry ●...
  • Page 14: Contraindication By Instruments And Other Combined Devices

    1 General information Dentsply Sirona 1.11 Contraindication by instruments and other combined devices Instructions for Use Treatment Center Axano Pure 1.11 Contraindication by instruments and other combined devices Contraindications for the following tools, which can be operated on the product (combined devices), are to be taken from the separate documentation provided for this purpose: ●...
  • Page 15: Obligation To Notify Authorities

    Dentsply Sirona 1 General information Instructions for Use Treatment Center Axano Pure 1.14 Obligation to notify authorities with the legal and normative standards for infection prevention in dentistry. ● Cleaning personnel The cleaning personnel carry out the general cleaning and disinfection in the dental practice according to a work schedule.
  • Page 16: Safety Notices

    2 Safety notices Dentsply Sirona 2.1 Identification of danger levels Instructions for Use Treatment Center Axano Pure Safety notices Identification of danger levels To prevent personal injury and material damage, please observe the warning and safety information provided in these operating instructions.
  • Page 17: On-Site Installation

    Risk of infection and poisoning Poor water quality in the public water supply can lead to infections and poisoning. Dentsply Sirona cannot influence the water quality. The operator is responsible for water quality. Test the water quality from the treatment center at regular intervals and after it has not been used for >1 week, see "Microbiological water test" ...
  • Page 18: Maintenance And Repair

    2 Safety notices Dentsply Sirona 2.7 Maintenance and repair Instructions for Use Treatment Center Axano Pure the transport of drinking water. It can be connected directly to the public drinking water supply. The treatment center bears the DVGW label. Always adhere to the national requirements with regard to connecting treatment centers to the public drinking water supply.
  • Page 19: Patient Chair

    No restrictions apply when using salt blasting devices in connection with Dentsply Sirona treatment centers. However, in such cases, make sure that the system is subsequently flushed with an adequate amount of water.
  • Page 20: Care And Cleaning Instructions For The Practice Team

    ● connections are encrypted with SSL/TLS and are only established with a valid certificate ● connections must be allowed by the user on the Smart Touch ● only software signed by Dentsply Sirona can be installed on the treatment center Despite these measures, we recommend: ●...
  • Page 21: Electrostatic Discharge

    Dentsply Sirona 2 Safety notices Instructions for Use Treatment Center Axano Pure 2.18 Electrostatic discharge vicinity of the treatment center. Non-compliance can lead to a reduction in the performance features of the treatment center. Sivision View and USB interface The presence of electromagnetic interference in the vicinity of the treatment center may cause image interference and interruptions in the data transmission via the USB interface to the PC.
  • Page 22: Unit Description

    3 Unit description Dentsply Sirona 3.1 Operating principle Instructions for Use Treatment Center Axano Pure Unit description Operating principle The device is designed to bring a patient into an ergonomic and comfortable position to perform dental treatment procedures. Patient chair...
  • Page 23 Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.1 Operating principle Operating light An operating light providing illumination to the oral cavity can be switched via the Smart Touch or is operated automatically in accordance to preset adjustments in connection with the preprogrammed chair positions.
  • Page 24: Standards/Approvals

    3 Unit description Dentsply Sirona 3.2 Standards/Approvals Instructions for Use Treatment Center Axano Pure Standards/Approvals The Axano Pure treatment center complies with the following standards, among others: ● IEC 60601-1 (electrical, mechanical, and software safety) ● IEC 80601-2-60 (safety of dental equipment) ●...
  • Page 25 Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.2 Standards/Approvals the chapter entitled "Connection to the public drinking water system"  [→ 17]. The wireless modules in the wireless foot control and in the treatment center meet the requirements of the RED Directive 2014/53/EU.
  • Page 26: Technical Data

    3 Unit description Dentsply Sirona 3.3 Technical data Instructions for Use Treatment Center Axano Pure Technical data Model designation: Treatment center Axano Pure Power connection: 100 – 240 V AC ± 10  % 50/60 Hz Permanently connected unit Nominal current: 6 – 2.5 A plus max. 6 A for external devices Type of ground connection: TN-C-S system or TN-S system (acc.
  • Page 27 Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.3 Technical data 1060 WARNING Operating conditions Ambient temperature: 10 °C – 40 °C (50 °F – 104 °F) Relative humidity: 30% – 85%, no condensation Air pressure: 700 hPa – 1060 hPa Installation location: ≤ 3000 m above sea level Operation is not permitted in mobile ve- hicles.
  • Page 28: System Overview

    3 Unit description Dentsply Sirona 3.4 System overview Instructions for Use Treatment Center Axano Pure System overview The treatment center comprises the following main components: Patient chair Motor-driven headrest or double articulating headrest Smart Control (shown) or Classic foot control...
  • Page 29: Patient Chair

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.5 Patient chair Patient chair The patient chair features a variety of adjustment options to optimally adjust the patient's position to the treatment. Motor-driven headrest (shown here) or double articulating...
  • Page 30 3 Unit description Dentsply Sirona 3.5 Patient chair Instructions for Use Treatment Center Axano Pure The patient chair can be equipped with lounge upholstery featuring enhanced comfort and double seams. The lounge upholstery has no footrest; the padded surface is continuous.
  • Page 31: Motor-Driven Headrest

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.6 Motor-driven headrest Motor-driven headrest The headrest allows for the following adjustment options: ● Motorized extension/retraction to adapt to the patient's height ● Motor-driven tilting for maxillary/mandibular treatment ●...
  • Page 32 3 Unit description Dentsply Sirona 3.7 Double-jointed head support Instructions for Use Treatment Center Axano Pure Head pad Rotary joints Unlock button (one-sided) Support bar for adjustment to patient height For details, see "Adjusting the double articulating headrest."  [→ 69] 68 29 514 D3685...
  • Page 33: Foot Control

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.8 Foot control Foot control The foot control enables hands-free control of the treatment center. Via the integrated cursor control, virtually all functions of the treatment center can also be controlled via the foot control as an alternative to the Smart Touch.
  • Page 34: Classic Foot Control

    3 Unit description Dentsply Sirona 3.8 Foot control Instructions for Use Treatment Center Axano Pure 3.8.2 Classic foot control The foot control has a left and right key (B) and (D). The 4-way control plate (C) is activated by moving it to the left/right and up/down.
  • Page 35: Dentist Element

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.9 Dentist element Dentist element All functions of the treatment center can be controlled via the Smart Touch control panel on the dentist element. The dentist element is moved along the manual sliding track.
  • Page 36: Instrument Positions

    3 Unit description Dentsply Sirona 3.9 Dentist element Instructions for Use Treatment Center Axano Pure 3.9.1 Instrument positions The following instrument position assignments are possible on the dentist element: Position 1 Position 2 Position 3 Position 4 Position 5 Position 6 Sprayvit BL E motor...
  • Page 37: Smart Touch User Interface

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.9 Dentist element 3.9.2 Smart Touch user interface Axano EXCA PREP Standby switch Dialog area Shows the main dialogs and their extensions. The functions keys displayed depend on the operating context.
  • Page 38: Assistant Element

    3 Unit description Dentsply Sirona 3.10 Assistant element Instructions for Use Treatment Center Axano Pure 3.10 Assistant element The functional scope of the assistant element is adapted to the dental assistant's field of activity. It can, however, also be positioned so as to enable unassisted treatment by the dentist.
  • Page 39: User Interface

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.10 Assistant element On the underside of the assistant element, a medical power supply for Type BF applied parts is available (12  V  DC, max. 1  A ), e.g. for supplying light for a suction handpiece.
  • Page 40 3 Unit description Dentsply Sirona 3.10 Assistant element Instructions for Use Treatment Center Axano Pure Bell Freely available relay 240 VAC, 6 A (connected by the service technician) This function can be preset as a button or a switch in the setup dialog, see "Maintenance/Device settings ...
  • Page 41: Water Unit

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.11 Water unit 3.11 Water unit The water unit is equipped with a disinfection system. In normal operation, this will automatically inoculate the water that comes in contact with the patient (also called treatment water) with a disinfectant for the water lines.
  • Page 42 3 Unit description Dentsply Sirona 3.11 Water unit Instructions for Use Treatment Center Axano Pure Suction hose connection for assistant element Swiveling cuspidor Tumbler filling (depicted) or tumbler filling with automatic sen- sor control for automatic tumbler filling Cover of the storage tank for the disinfectant for the water...
  • Page 43: External Device Connection

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.12 External device connection 3.12 External device connection DANGER External medical additional devices can be connected to the external device connection. They must comply with the requirements of IEC 60601-1.
  • Page 44 3 Unit description Dentsply Sirona 3.12 External device connection Instructions for Use Treatment Center Axano Pure Fuse for IEC socket (6.3 A slow-blow) IEC socket with power supply (max. 6 A) Pressure Flow rate Water 2 to 3 bar max. 300 ml/min 4.7 ±...
  • Page 45: Digital Services

    Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.13 Digital services 3.13 Digital services We recommend connecting the treatment center to the internet via the practice network. Only this ensures that you can utilize the full potential of your new treatment center.
  • Page 46: Combination With Other Devices

    3 Unit description Dentsply Sirona 3.14 Combination with other devices Instructions for Use Treatment Center Axano Pure 3.14 Combination with other devices Straight and contra-angle handpieces Straight and contra-angle handpieces are medical devices which are connected to a standardized interface.
  • Page 47 Dentsply Sirona 3 Unit description Instructions for Use Treatment Center Axano Pure 3.14 Combination with other devices Axano, Axano Pure, Teneo, Sinius, and Intego treatment centers. The LEDview Plus is attached to the lamp support tube of the treatment center. It is operated via the user interface of the treatment center and a proximity sensor.
  • Page 48: Consumption Of Water, Dentosept, And The Agent For Suction Hose And Cuspidor Cleaning

    3 Unit description Dentsply Sirona 3.15 Consumption of water, Dentosept, and the agent for suction hose and cuspidor cleaning Instructions for Use Treatment Center Axano Pure 3.15 Consumption of water, Dentosept, and the agent for suction hose and cuspidor cleaning...
  • Page 49: Operation

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.1 Starting up the treatment center Operation Starting up the treatment center 4.1.1 Switching the treatment center on/off The treatment center is equipped with a standby system for enhanced convenience when switching it on and off at the dentist element.
  • Page 50: Clinic Mode

    4 Operation Dentsply Sirona 4.1 Starting up the treatment center Instructions for Use Treatment Center Axano Pure ü The power switch is turned on. ➢ Press the standby switch on the dentist element. Ä The LED of the standby switch lights up on the dentist element.
  • Page 51: Initial Startup

    For further details, please refer to the section "Usage data  [→ 12]". Registering the treatment center ➢ Have your treatment center registered with Dentsply Sirona by your service technician. Please note that the full extent of functions of the service software,...
  • Page 52: Logging User In/Out

    4 Operation Dentsply Sirona 4.1 Starting up the treatment center Instructions for Use Treatment Center Axano Pure 3. Tap on the blue text "Add user". 4. Enter your name and email address. A keyboard appears when you tap on the fields. You may optionally protect your user profile with a password.
  • Page 53 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.1 Starting up the treatment center Change user profile 1. Tap on the name of the active user profile in the header. Ä The user selection opens. 2. Select the desired user profile. Touch the name of the desired profile.
  • Page 54: Operating Concept Of The Smart Touch

    4 Operation Dentsply Sirona 4.2 Operating concept of the Smart Touch Instructions for Use Treatment Center Axano Pure Operating concept of the Smart Touch On the Smart Touch user interface, the desired functions can be activated either by tapping them with your finger or via the cursor control of the foot control.
  • Page 55: Main Dialogs

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.2 Operating concept of the Smart Touch When moving the chair into position, the selected program key has a gray background. The key is displayed in orange when the chair position is reached.
  • Page 56: Expanded Dialogs

    4 Operation Dentsply Sirona 4.2 Operating concept of the Smart Touch Instructions for Use Treatment Center Axano Pure Customizing the Smart Touch dialogs The dialogs can be customized for each user, e.g. function keys can be removed from dialogs, added to them, or positioned in a different place, see "User interface ...
  • Page 57 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.2 Operating concept of the Smart Touch Footer The dialog change keys on the lower edge of the Smart Touch can be used to switch between the main dialogs: ●...
  • Page 58: Foot Control

    4 Operation Dentsply Sirona 4.3 Foot control Instructions for Use Treatment Center Axano Pure Foot control The treatment center can be operated using a wireless foot control or a foot control with a cable connection. 4.3.1 Wireless foot control Technical data of the wireless module, see Foot control wireless interface" ...
  • Page 59 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.3 Foot control Middle button and swivel buttons of the Smart Control If the cursor control is switched on, it is operated via the middle button and swivel buttons, see "Using the cursor control"  [→ 60].
  • Page 60: Using The Cursor Control

    4 Operation Dentsply Sirona 4.3 Foot control Instructions for Use Treatment Center Axano Pure 4.3.3 Using the cursor control 4.3.3.1 Functionality Cursor control as an alternative mode of operation The Smart Touch can also be operated hands-free via the foot control.
  • Page 61: Using Cursor Control With Comfort Foot Control

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.3 Foot control Cursor path The cursor path runs along the pairs of keys, generally moving from top to bottom and from left to right, usually in multiple loops. The cursor path can be traversed either in a forward or a reverse direction.
  • Page 62 4 Operation Dentsply Sirona 4.3 Foot control Instructions for Use Treatment Center Axano Pure ➢ Press and hold the middle button (> 600 ms). Ä The red cursor moves back to its starting position. In the start and Sivision dialog, this is the position after the blue pair of keys, in the instrument dialog the quick setting keys.
  • Page 63: Using Cursor Control With Classic Foot Control

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.3 Foot control 3. Confirm your selection by pressing the middle button. Ä The selected main dialog is displayed on the Smart Touch. 4.3.3.3 Using cursor control with Classic foot control Moving the cursor ➢...
  • Page 64 4 Operation Dentsply Sirona 4.3 Foot control Instructions for Use Treatment Center Axano Pure Ä The dialog change keys have a red border. 2. Slide the 4-way foot control plate to the right or left to select the desired main dialog.
  • Page 65: Patient Chair

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.4 Patient chair Patient chair 4.4.1 Safety instructions WARNING The free space under the patient couch and up to the water unit can be decreased due to chair movements.
  • Page 66: Safety Stop

    4 Operation Dentsply Sirona 4.4 Patient chair Instructions for Use Treatment Center Axano Pure 4.4.2 Safety stop The treatment center is equipped with various safety stops to prevent crushing and damage: ● Cuspidor ● Lift frame ● Assistant element ● Rear cover, right/left ●...
  • Page 67: Triggering An Immediate Movement Stop

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.4 Patient chair 4.4.3 Triggering an immediate movement stop WARNING You can stop the movement of the chair to a programmed position as follows: ➢ Tap on a key for the patient chair on the Smart Touch, a key to tilt the motorized headrest.
  • Page 68: Armrests

    4 Operation Dentsply Sirona 4.4 Patient chair Instructions for Use Treatment Center Axano Pure 4.4.4 Armrests Armrests are available for the patient chair. The armrest nest to the dentist element can be swiveled forward to make it easier for patients to enter and exit the chair.
  • Page 69: Adjusting The Double Articulating Headrest

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.4 Patient chair 4.4.6 Adjusting the double articulating headrest The double articulating headrest is equipped with two rotary joints. They facilitate manual adjustment of head inclination for maxillary/mandibular treatments. The headrest extension can be pushed in or pulled out manually to adjust it to the height of the patient.
  • Page 70: Moving The Patient Chair To The Entry/Exit Position

    4 Operation Dentsply Sirona 4.4 Patient chair Instructions for Use Treatment Center Axano Pure 4.4.7.1 Moving the patient chair to the entry/exit position The following functions are triggered to facilitate patient entry and exit in the entry/exit position: ● The patient chair moves to an upright position ●...
  • Page 71: Moving The Patient Chair To The Mouth Rinsing Position

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.4 Patient chair 4.4.7.2 Moving the patient chair to the mouth rinsing position The following functions are triggered in the mouth rinsing position: ● The operating light switches off ●...
  • Page 72: Moving The Patient Chair To A Treatment Position

    4 Operation Dentsply Sirona 4.4 Patient chair Instructions for Use Treatment Center Axano Pure ü The patient chair can be in any treatment position. 1. Tap on the S key on the Smart Touch, or press the S key on the user interface of the assistant element, or press the left button of the foot control (with all instruments in place in their holders).
  • Page 73: Moving The Chair Manually

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.4 Patient chair 4.4.8 Moving the chair manually IMPORTANT Chair movements with cuspidor swiveled in Manual cuspidor: With the cuspidor swiveled in, the chair cannot be moved. This prevents the patient from colliding with the cuspidor.
  • Page 74: Adjusting The Chair Height

    4 Operation Dentsply Sirona 4.4 Patient chair Instructions for Use Treatment Center Axano Pure Via the 4-way foot switch In the setup dialog, the ErgoMotion function can be set so it can also be operated via the 4-way foot switch, see "Settings/Operation/4-way foot switch ...
  • Page 75 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.4 Patient chair Movement via Smart Touch ➢ If a motor-driven headrest is available: Touch the Tilt headrest , Move headrest in/out , ErgoMotion , or Chair height adjustment keys.
  • Page 76: Configuring Chair Programs And Shock Positioning

    4 Operation Dentsply Sirona 4.4 Patient chair Instructions for Use Treatment Center Axano Pure 4.4.9 Configuring chair programs and shock positioning Chair programs The six factory set chair programs: ● Mouth rinsing position S ● Entry/exit position 0 ● 1 and 2 ●...
  • Page 77: Treatment While Standing

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.4 Patient chair 4.4.10 Treatment while standing For treatment while standing, a second level can be activated for the chair programs 1 to 4. ü The start dialog is displayed.
  • Page 78 4 Operation Dentsply Sirona 4.4 Patient chair Instructions for Use Treatment Center Axano Pure Ä If chair positions are already saved in Sidexis 4 for the registered patient, a "P" will be shown on the respective program keys. 2. Touch key S, 0, 1, 2 or, if applicable, 3, 4 briefly (< 2 s).
  • Page 79: Dentist Element

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element Dentist element 4.5.1 Safety notices WARNING The free spaces near the patient can be reduced when the dentist element is moved. Parts of the patient's or user's body may be pinched or crushed.
  • Page 80: Positioning The Dentist Element

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.4 Positioning the dentist element The dentist element can be moved along the entire length of the patient chair. In combination with the rotary joints on the support arm, the dentist element can be perfectly adjusted to suit any treatment situation.
  • Page 81: Fixed Keys On The Dentist Element

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.5 Fixed keys on the dentist element Axano The keys to be displayed in the fixed key area can be set in the setup dialog. The position of the fixed keys can be changed. See "User interface ...
  • Page 82: Composite Function

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.5.5 Composite function The composite function delays the curing of composite materials. ➢ Press the Composite function fixed key. Ä When the composite function is switched on, the key is marked orange on the Smart Touch.
  • Page 83: Freely Selectable Function

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element Two displays are available for timers, as a single timer and as a multitimer. The display can be configured as a widget, see User interface  [→ 183]/widget.
  • Page 84: Placing The Instruments In Their Holders

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.6 Placing the instruments in their holders Automatic opening of instrument dialogs The instrument dialog that corresponds to the instrument taken from the holder, with the exception of the Sprayvit and SmartLite Pro, is automatically displayed on the Smart Touch.
  • Page 85: Saving Settings

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.7 Saving settings Quick setting keys Instrument settings can be saved and accessed on the quick setting keys individually for each user profile. In the Setup dialog, you can set whether the settings coolant,...
  • Page 86 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 68 29 514 D3685 D3685.201.02.01.02    2023-12...
  • Page 87: General Instrument Functions

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.8 General instrument functions Settings for the coolant, instrument light, and foot control can be made in the expanded instrument program of the instrument removed from the holder.
  • Page 88: Setting The Amount Of Spray Water

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure ü A turbine or motor is removed from the holder. ➢ Tap on the Chip blower key on the Smart Touch. ➢ Press the right button of the foot control.
  • Page 89: Switching The Instrument Light On/Off And Adjusting The Brightness

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.8.5 Switching the instrument light on/off and adjusting the brightness Switch instrument light on/off ü The instrument dialog of the removed turbine, motor, or ultrasonic handpiece is displayed on the Smart Touch.
  • Page 90 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure Meaning: Caution, risk for fingers in contact with moving parts Close cover (A) before operating the NaCl pump. WARNING Fine objects can enter the mechanism of the NaCl pump via openings.
  • Page 91: Setting The Nacl Flow Rate

    Ultrasonic scaler tips of third-party manufacturers do not always have a sufficient flow rate in conjunction with the NaCl function. ➢ Use only ultrasonic scaler tips from Dentsply Sirona. See also the SiroSonic TL instructions for use. 68 29 514 D3685...
  • Page 92: Sprayvit Multifunctional Syringe

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.9 Sprayvit multifunctional syringe The Sprayvit multifunctional syringe is used for dental treatment with air and water. The Sprayvit multifunctional syringe is also used to illuminate the preparation field.
  • Page 93: Connecting The Instrument Hose

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element WARNING A small metal tube protrudes at the tip of the nozzle. Risk of injury and contamination! ➢ Avoid contact with the tip of the nozzle. Do not attempt to handle with the nozzle attached.
  • Page 94: Applying Air, Water, Or Spray

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.9.7 Applying air, water, or spray In the standard design, the water button for the Sprayvit multifunctional syringe is located on the right. A Sprayvit with inverted media (water on the left) is available as an option.
  • Page 95 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element Ä In the start dialog , the Sprayvit instrument light is displayed; in the expanded start dialog , the Sprayvit water tempering key is also displayed.
  • Page 96: Turbine / Air Motor / Other Air-Driven Instruments

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.10 Turbine / air motor / other air-driven instruments WARNING Excessive speed can cause damage to the teeth. ➢ Note speed limit for the tool. The permissible speed is indicated in the tool documentation.
  • Page 97: Motor

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.11 Motor The motor drives rotating and oscillating straight and contra-angle handpieces. In addition, motorized endometric measurement is supported. The motors meet the ISO 14457 standard. 4.5.11.1...
  • Page 98 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure ISO adapter (left) and Basic Apex adapter (right) BL ISO F motor Handpiece holder O-ring 8.4 x 0.7 Product labeling BL ISO F motor sleeve Coupling connection (green / 3 guide lugs) The BL ISO F motor is equipped with an ISO interface.
  • Page 99: Product Labeling

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.11.2 Product labeling The motors or the motor sleeves and the adapters are labeled with the following information: Manufacturer and Sterilizable at product name 135 °C (275 °F)
  • Page 100: Safety Instructions

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure BL E and BL Implant E BL ISO F Spray air flow in L/min > 1.5 – Spray water pressure in bar 2 ±0.2 – Spray water flow in ml/min >...
  • Page 101: Connecting The Instrument Hose

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element NOTICE Protecting the motor shaft/axis A bent motor shaft on the BL E motor causes irregular operating noises or strong vibrations. This can damage instruments. ➢ Do not let the motor fall to the floor.
  • Page 102: Setting The Speed

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure Attach the instrument to the BL ISO F motor and BL Implant E motor or adapter ü The motor is at a standstill. 1. Align the nib (A) of the instrument with the groove (B) of the motor.
  • Page 103: Changing Direction Of Rotation

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element ➢ Position the cursor on the rpm field by pressing and holding the middle button. Then press the swivel buttons. The size of the increments depends on the speed range set: From 90 to 100 rpm = 10 rpm increments...
  • Page 104 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure ➢ Position the red cursor on the CCW rotation key by pressing the middle button briefly. Then, depending on the position of the key, press the left or right swivel button of the foot control. The cursor automatically moves back to the starting position (quick setting keys).
  • Page 105: Sirosonic Tl Ultrasonic Handpiece

    Ultrasonic tips from other manufacturers do not guarantee safe operation and can lead to harm. ➢ Use only original Dentsply Sirona parts or parts/tips approved by Dentsply Sirona. Safe operation is not guaranteed for parts/tips that have not been approved by Dentsply Sirona and the handpiece may be damaged.
  • Page 106 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure Changing the intensity 1. Set the desired intensity with the slider. ➢ Tap on the - or + keys in the Ultrasonic intensity field. Ä The intensity value is increased or reduced.
  • Page 107: External Hf Surgical Device

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.13 External HF surgical device External HF surgery units may interfere with the treatment center and patient monitor. The treatment center should therefore be protected against interference caused by the external HF surgery unit whenever the latter is used.
  • Page 108: Endodontic And Implantology Treatments

    4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.14 Endodontic and implantology treatments The treatment functions support endodontic and implantology treatments. A selection of the most common file and implantology systems with the manufacturer's recommended speed and torque values is stored in a database.
  • Page 109 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element Standard endodontics dialog Advanced endodontics dialog 2. Take the motor that you wish to assign to the endodontic treatment from the instrument holder. Ä Next to the symbol of the endodontics dialog change key, a number is displayed that shows the instrument position of the motor removed.
  • Page 110 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.14.1.2 Calibrating the burr drive WARNING Calibration must be performed each time you start treatment and each time you change or lubricate the contra-angle handpieces. Recalibration is not necessary when a file is changed.
  • Page 111 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element ➢ The file selection is the starting position for the red cursor. Press the left or right swivel button of the foot control. Ä The selected file is marked.
  • Page 112 Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure ü The desired file is selected in the advanced endodontics dialog . ➢ Use the – and + keys to set the speed and torque of the contra- angle handpiece.
  • Page 113 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element After the electric motor is started with the foot control, an audible warning signal sounds if counterclockwise rotation is activated. ü The endodontics dialog is displayed on the Smart Touch.
  • Page 114 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.14.1.10.1 Preparing to use the ApexLocator Apex adapter, mucosal electrode and file clamp The mucosal electrode and the file clamp are connected to the socket below the dentist element using the apex adapter.
  • Page 115 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element For endodontic treatments with the ApexLocator, the Dentsply Sirona Endo 6:1 or Endo 6 L contra-angle handpieces are required for the treatment function. WARNING Pull the silicone isolation sleeve (A) over the contra-angle handpiece and wear insulating gloves to prevent faulty measurements due to leakage current.
  • Page 116 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure Colored regions of the distance display Blue region The tip of the root canal file is near the apex. Green region The tip of the root canal file has reached the physiological apex.
  • Page 117 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.14.1.10.3 Apex and endodontics beeps Apex beeps In addition to the graphical distance display on the Smart Touch, the position of the file in the root canal can also be indicated as an acoustic signal.
  • Page 118 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure The sounding of a signal when 75% of the set torque value is reached can be switched on/off in the expanded endodontics dialog, see "Switch torque signal on/off"  [→ 112].
  • Page 119 [→ 115]. Ä The maximum root canal depth reached is displayed by two green arrowheads. 4.5.14.1.11 Switching the instrument light on/off and adjusting the brightness The Dentsply Sirona ENDO 6 L handpiece is equipped with a fiber optic. 68 29 514 D3685 D3685.201.02.01.02    2023-12...
  • Page 120 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure ü The expanded endodontics dialog is displayed on the Smart Touch. 1. Use the – and + keys to adjust the brightness of the instrument light.
  • Page 121 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.14.1.12 Managing file systems 4.5.14.1.12.1 Opening the endodontic management dialog ü The endodontics dialog is displayed on the Smart Touch. 1. Tap on the Select file system key.
  • Page 122 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure You can add your own files to your system. 1. Tap on the Create instrument key. 2. Assign a name to your individual file under Details and set the desired values.
  • Page 123: Implantology

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 4.5.14.2 Implantology WARNING Only the Bien-Air CA 20:1 L implant contra-angle handpiece may be used for the implantology treatment. Instruments from other manufacturers can lead to malfunctions or be incorrectly calibrated.
  • Page 124 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 3. Tap on the Cal key. Ä The button flashes orange. During calibration, increasing speeds are set on the motor. When the bur drive is calibrated, the key is permanently marked orange.
  • Page 125 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element ü The desired implant instrument is selected in the implantology dialog . ➢ Use the – and + keys to set the speed and torque of the contra- angle handpiece.
  • Page 126 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure ➢ Press the right button of the foot control. Ä The key is marked orange when CCW rotation is activated. 4.5.14.2.8 Switching the torque reverse function on/off...
  • Page 127 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element 2. Use the arrow keys to set the flow rate of the NaCl pump. Ä The selected setting is displayed on the upper edge of the Smart Touch.
  • Page 128 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.14.2.12 Managing implantology systems 4.5.14.2.12.1 Opening the implantology management dialog ü The implantology dialog is displayed on the Smart Touch. 1. Tap on the Select implantology system key.
  • Page 129 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.5 Dentist element You can add your own implant instruments to your system. 1. Tap on the Create instrument key. 2. Assign a name to your individual implant instrument under Details and set the desired values.
  • Page 130 4 Operation Dentsply Sirona 4.5 Dentist element Instructions for Use Treatment Center Axano Pure 4.5.14.2.13 Sivision View "Implant terminal" The most important parameters for implantology treatment can also be displayed on the patient monitor. The current speed and torque values measured by the control electronics are also sent to the monitor.
  • Page 131: Assistant Element

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.6 Assistant element Assistant element 4.6.1 Maximum load capacity WARNING The maximum load capacity of the assistant element is 1.5 kg (3.3 lbs.). Additionally, a skid-proof silicone mat can be placed on it. The silicone mat reduces the likelihood that instruments and treatment accessories can slide around.
  • Page 132: Fixed Keys On The Assistant Element

    4 Operation Dentsply Sirona 4.6 Assistant element Instructions for Use Treatment Center Axano Pure 4.6.3 Fixed keys on the assistant element 4.6.3.1 Tumbler filling If your treatment center is equipped with the tumbler filling option with automatic sensor control, see "Tumbler filling with automatic sensor control" ...
  • Page 133: Chair Programs

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.6 Assistant element ➢ Press the Composite function fixed key. Ä If the composite function is switched on, the LED of the fixed key lights up on the assistant element. The key is marked orange on the Smart Touch.
  • Page 134: Suction Handpieces

    4 Operation Dentsply Sirona 4.6 Assistant element Instructions for Use Treatment Center Axano Pure 4.6.4 Suction handpieces The assistant element can be equipped with a maximum of three suction handpieces. Suction cannulas with a diameter of 11 mm and 6 mm can be used.
  • Page 135: Smartlite Pro Curing Light

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.6 Assistant element Saliva ejector A curved cannula can be placed in the corner of the mouth to aspirate saliva. The cannulas can be ordered from a specialized dealer.
  • Page 136: Water Unit

    4 Operation Dentsply Sirona 4.7 Water unit Instructions for Use Treatment Center Axano Pure Water unit 4.7.1 Swiveling the cuspidor bowl The cuspidor on the water unit can be swiveled manually or automatically (optional) by approx. 110 mm toward the patient chair.
  • Page 137: Adjusting The Water Amount For Flushing

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.7 Water unit If the tumbler is removed before the preset filling level has been reached, the water flow stops immediately. ü The tumbler may not be transparent. ü The tumbler must always be placed in the center of the area provided for it.
  • Page 138 4 Operation Dentsply Sirona 4.7 Water unit Instructions for Use Treatment Center Axano Pure WARNING Microorganisms can multiply in the water. These microorganisms could be a health risk. ➢ Mix fresh water for the self-contained water supply every day. At the end of the work day, the remaining water must be flushed out of the disinfectant tank using the tumbler filling function.
  • Page 139 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.7 Water unit Ä The disinfectant tank is drained. The switch to self-contained water supply is indicated by four blue dots on the touchscreen. Ä If there is still undiluted disinfectant for the water lines in the tank, it is rinsed down the drain.
  • Page 140 4 Operation Dentsply Sirona 4.7 Water unit Instructions for Use Treatment Center Axano Pure Resetting the disinfection system to operation with public drinking water and the disinfectant for the water lines The disinfection system is reset to public drinking water mode in the same manner as switching to self-contained water supply.
  • Page 141: Tray

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.8 Tray Tray Maximum load capacity WARNING The maximum load is 1 kg (2.2 lbs). Removing the tray 1. Hold the tray tightly. 2. Open the lock by swiveling the lever downwards.
  • Page 142: Cup Holder

    4 Operation Dentsply Sirona 4.9 Cup holder Instructions for Use Treatment Center Axano Pure Removing the brake knob If the brake knob of the support arm is considered annoying, this can be removed from the support arm. 1. Gently loosen the brake knob on safety bolt (B).
  • Page 143: Smart Device Holder

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.10 Smart device holder 4.10 Smart device holder The dentist element can be equipped with a holder for a smart device. WARNING It must be possible to disinfect the smart device.
  • Page 144: Panoramic X-Ray Viewer

    4 Operation Dentsply Sirona 4.11 Panoramic X-ray viewer Instructions for Use Treatment Center Axano Pure 4.11 Panoramic X-ray viewer The dentist element can be equipped with an X-ray viewer. 4.11.1 Switching the X-ray viewer on/off ü The expanded start dialog is displayed on the Smart Touch.
  • Page 145: Operating Light

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.12 Operating light 4.12 Operating light The operating light is mounted on a multi-joint support arm. It can be easily adjusted to the operating field using the handles. Brakes in the support arm hold the operating light in the position to which it has been adjusted.
  • Page 146 4 Operation Dentsply Sirona 4.12 Operating light Instructions for Use Treatment Center Axano Pure In the setup dialog, the composite function can be set so it can be switched on/off via the 4-way foot switch, see "Settings/Operation/4-way foot switch  [→ 172]". The spray aspirator must be put in the holder if it has been set in the setup dialog that the suction flow can be switched on/off via the 4-way foot switch.
  • Page 147: X-Ray Tube Unit

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.13 X-ray tube unit 4.13 X-ray tube unit The treatment center can be equipped with the Heliodent Plus X-ray tube unit. The X-ray tube unit is attached to the light post of the treatment center using a multi-jointed support arm that can be adjusted both horizontally and vertically.
  • Page 148: Sivision Video System

    4 Operation Dentsply Sirona 4.14 Sivision video system Instructions for Use Treatment Center Axano Pure 4.14 Sivision video system The integrated Sivision video system enables intraoral and extraoral images to be made. The SiroCam AF / AF+ intraoral camera generates digital image data that can be transmitted via a USB 2.0 port (high-...
  • Page 149: Sirocam Af/Af+ Intraoral Camera

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.14 Sivision video system Attachment versions Monitor on lamp support tube The monitor can be rotated and swiveled. Monitor on tray arm The monitor can be rotated and swiveled.
  • Page 150: Connecting The Intraoral Camera

    4 Operation Dentsply Sirona 4.14 Sivision video system Instructions for Use Treatment Center Axano Pure Lens window (covered) Configurable camera key for Delete last image, Focusing, or Without function, see in the "Settings/Instruments  [→ 167]/ Camera button“ setup Connection cable 4.14.2.3 Connecting the intraoral camera The intraoral camera is connected on the bottom of the dentist element.
  • Page 151 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.14 Sivision video system Switch camera light on/off ➢ Tap on the Camera light key on the Smart Touch. ➢ Press the right swivel button of the Smart Control.
  • Page 152 4 Operation Dentsply Sirona 4.14 Sivision video system Instructions for Use Treatment Center Axano Pure Switching autofocus and focus frame on/off When the key is not marked orange, the image is automatically refocused after each exposure. This takes a short moment.
  • Page 153 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.14 Sivision video system You can also use the right foot control button to switch between the views. Deletes the currently marked image. In the Setup dialog, you can set that the current images can also be deleted by the button on the camera, see "Settings/Instruments ...
  • Page 154: Technical Camera Data

    4 Operation Dentsply Sirona 4.14 Sivision video system Instructions for Use Treatment Center Axano Pure Exporting image to Sidexis 4 The key exports the current photo to Sidexis 4. The images can be assigned to a patient and saved in the patient database via file import, see "Sidexis 4"...
  • Page 155: Operation With A Pc

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.15 Operation with a PC 4.15 Operation with a PC The treatment center can be connected to a PC via an Ethernet cable. The network connection to the treatment center can be configured in the "Maintenance/Network ...
  • Page 156 4 Operation Dentsply Sirona 4.15 Operation with a PC Instructions for Use Treatment Center Axano Pure The touchpad is behind a hygienic glass surface. Comfortable operation of the mouse cursor depends on the quality of the medical gloves. Not all gloves move equally well across the glass surface. The best results are achieved with gloves that are latex-free and fit tightly.
  • Page 157: Media Player

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.15 Operation with a PC Smooth image To mitigate high-contrast or high-interference effects in images, the contrast between neighboring pixels is reduced or averaged. The overall sharpness of the image is reduced.
  • Page 158 4 Operation Dentsply Sirona 4.15 Operation with a PC Instructions for Use Treatment Center Axano Pure Stop playback Forward/Back Adjust volume Mute Enlarge/reduce view Show previous/next /page Adjust to page size Adjust to page width Close file 68 29 514 D3685...
  • Page 159: Usb Port

    There is a risk of electric shock. ➢ Only type BF medical devices and applied parts according to IEC 60601-1, e.g., the Dentsply Sirona Xios XG USB module intraoral X-ray system, may be connected to the USB interface. ➢ Do not connect a USB hub or extension cables to the USB interface.
  • Page 160: Configuration Of The Treatment Center (Setup)

    4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) In the setup screen, the treatment center can be individually set up and configured. 4.16.1 Opening the setup dialog ü...
  • Page 161: Notifications

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) "Notifications" dialog 2. The setup is divided into 6 dialogs. Select the desired dialog on the left side of the Smart Touch.
  • Page 162: User Profiles

    4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.3 User profiles Multiple users can use the treatment center with their own individual treatment and operating settings. The following is stored in the user profiles: ●...
  • Page 163: Bluetooth

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Sorting user names The following display sequences are available for selecting a user: ● Alphanumeric order by user name ● Sorting by last login Deleting user profile User profiles can be deleted completely.
  • Page 164: Settings

    Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure Ä The serial number of the wireless foot control is displayed under "New devices". It can be found on the rating plate of the foot control.
  • Page 165: General

    Registration is required for service offers such as firmware updates, product improvements, and expanded functions. Your registration data filed with Dentsply Sirona are displayed in the setup dialog and in the DS service software when you log in as a practice employee.
  • Page 166 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.5.1.1 Date & time Time format You can select either a 12-hour or 24-hour format. The 12-hour display is shown only in the header of the Smart Touch. The 24-hour format will still be used in the setup dialog.
  • Page 167: Treatment Chair

    Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Default display on patient monitor in exit position If the function is activated, the Dentsply Sirona logo is displayed when the chair program entry/exit position 0 is activated. 4.16.5.2...
  • Page 168 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure Expanded favorites storage Instrument settings can be saved and accessed individually on the quick setting keys for each user profile. One of two operating modes can be selected: ●...
  • Page 169 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Water temperature The temperature of the treatment water for the instruments on the dentist and assistant element can be set in 5 increments. Check the water temperature after filling a cup.
  • Page 170: Operation

    4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.5.4 Operation Cursor control The hands-free operation of the user interface via cursor control can be set as follows: ● off: No hands-free operation ●...
  • Page 171 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Display brightness The brightness of the Smart Touch will decrease after a few years of operation. This aging effect is normal and can be offset by adjusting the brightness.
  • Page 172 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.5.4.1 4-way foot switch 4-way foot switch horizontal and 4-way foot switch vertical You can select which function the 4-way foot switch triggers in the directions left/right and up/down: ●...
  • Page 173 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Suction stop for large suction devices If the treatment center is equipped with two spray aspirators, you can select in the setup dialog whether only one or both aspirators can be switched on/off with the 4-way foot switch or the star key on the assistant element.
  • Page 174 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.5.4.3 Sounds General volume The general device volume for non-critical system sounds such as moving up and down, confirmations, or notification sounds can be set in three levels.
  • Page 175: Operating Light

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) 4.16.5.5 Operating light Brightness The brightness of the operating light in normal mode (not composite) can be set in up to seven levels. If fewer than seven levels can be set, brightness has been limited in the setup dialog "Maintenance/Device...
  • Page 176: Hygiene

    4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.5.6 Hygiene 4.16.5.6.1 Tumbler Filling quantity The filling quantity of the tumbler can be set in increments. For tumbler filling with automatic sensor control, the water volume is regulated by the filling level, see "Tumbler filling with automatic sensor...
  • Page 177 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) 4.16.5.6.2 Cuspidor Automatically at S When the function is activated, cuspidor flushing is automatically switched on for the set time when the mouth rinsing position S is approached.
  • Page 178 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.5.6.3 Purge & autopurge Purge time For manual purging of the instruments, the purging time can be set between 20 and 180 seconds. The default setting is 20 seconds. For details on the purging process, please refer to section “Purging water...
  • Page 179 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Emptying of mixing tank After the autopurge process in the evening, the treatment center always empties the mixing tank of the disinfection system. When this function is activated, this is also done in the morning after the autopurge process.
  • Page 180 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.5.6.5 Suction hose cleaning Automatic cuspidor flushing with suction hose cleaning When the function is activated, chemical cuspidor cleaning is automatically carried out after every suction hose cleaning.
  • Page 181: User Settings

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Test of the rinsing tank sensor The service function can be used to check the function of the filling level sensor of the rinsing tank in the assistant element. The disinfectant tank is drained.
  • Page 182 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure and customization of the user interface are reset to the default settings preset in the factory. 68 29 514 D3685 D3685.201.02.01.02    2023-12...
  • Page 183: User Interface

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) 4.16.6 User interface To optimally support the workflow of each user, the keys in the main dialogs can be customized. Keys can be removed, added, or moved to a different position.
  • Page 184 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure Buttons Change the number of quick setting keys in the motor dialog You can add keys, remove them, or move them to a different position.
  • Page 185 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) information. At the moment, the widget for the timer and the multitimer can be selected here. 68 29 514 D3685 D3685.201.02.01.02    2023-12...
  • Page 186: Maintenance

    4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.7 Maintenance In the "Maintenance" area, relevant functions and settings are provided for your service technician. 4.16.7.1 Network Sivision Indicates the connection status of the treatment center to Sivision View.
  • Page 187: Device Settings

    Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) 4.16.7.2 Device settings Limitation of brightness The maximum brightness of the operating light in normal mode (not composite) can be limited to one of the 7 levels.
  • Page 188: Instruments

    For the LED lamp of each individual turbine, either the voltage or the current strength must be limited. The original Dentsply Sirona LED lamp is usually operated at 3.6 V. For lamps made by other manufacturers, the settings may have to be adjusted.
  • Page 189: Service Functions

    Uploading the device log When servicing, the log data of the treatment center can be sent to Dentsply Sirona for analysis and troubleshooting. They contain only setting and configuration data. To provide the data, the treatment center must have an internet connection, the remote service agreement must have been agreed to, and the treatment center must be registered via Siroforce Mobile.
  • Page 190 4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure Next maintenance date The treatment center must undergo an inspection and maintenance and a safety check at regular intervals, see "Maintenance by the service technician" ...
  • Page 191 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) Time cuspidor valve remains open If the treatment center is equipped with a cuspidor valve for central wet suction, the opening time of the valve can be extended between 0 and 2 seconds in 0.1-second increments.
  • Page 192: Software & Functions

    4 Operation Dentsply Sirona 4.16 Configuration of the treatment center (setup) Instructions for Use Treatment Center Axano Pure 4.16.7.5 Software & functions 4.16.7.5.1 Software Current version Display of software version currently on the treatment center The list shows whether newer or alternative software versions are available that can be installed on the treatment center.
  • Page 193 Dentsply Sirona 4 Operation Instructions for Use Treatment Center Axano Pure 4.16 Configuration of the treatment center (setup) 4.16.7.5.2 Functions Device UDI It is composed of the material number (REF) and the serial number of the treatment center. Activated functions Functions requiring a license can be activated using an activation code.
  • Page 194: Care, Cleaning And Maintenance By The Practice Team

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.1 Basics Instructions for Use Treatment Center Axano Pure Care, cleaning and maintenance by the practice team Basics 5.1.1 Intervals To maintain the value and safe functioning of your treatment center, it is necessary to have it regularly cared for, cleaned, and disinfected by the practice team.
  • Page 195 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.1 Basics ● Clean the filter in the suction hoses and disinfect the suction hoses  [→ 239] ● Disinfect handles and instrument holders and sterilize silicone mats ...
  • Page 196: General Processing Instructions

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.1 Basics Instructions for Use Treatment Center Axano Pure 5.1.2 General processing instructions The general reprocessing instructions apply to the treatment center in general provided there are no other product-specific reprocessing instructions in these instructions for use.
  • Page 197: Sterilization

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.1 Basics ISO 15883-1/-2, e.g. 93 °C (200 °F) and 10 min holding time. Refer to the instructions for use of the device for the respective use. After the process, check in good lighting (min.
  • Page 198: Care, Cleaning, And Disinfecting Agents

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.1 Basics Instructions for Use Treatment Center Axano Pure Cleaning and disinfection Sterilization Manual Automated Brush Wipe Washer- wrapped (only disinfec- wrapped cleaning) tor* Adapter  [→ 222] Ultrasonic handpiece and tips see separate Sirosonic TL instructions for use...
  • Page 199: Microbiological Water Test

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.1 Basics 5.1.5 Microbiological water test Perform the microbiological test of the water from the treatment center at regular intervals and after longer periods of disuse > 1 week; see "Media quality" ...
  • Page 200 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.1 Basics Instructions for Use Treatment Center Axano Pure 7. Count all bacterial colonies found on the surface of the tester. 68 29 514 D3685 D3685.201.02.01.02    2023-12...
  • Page 201: Surfaces

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.2 Surfaces Surfaces 5.2.1 Clean/disinfect plastic surfaces The plastic surfaces can be wiped with surface disinfectants. Observe the "General reprocessing instructions"  [→ 196].
  • Page 202: Disinfect User Interfaces

    WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! Surface disinfectants require application time, they dry off and leave streaks, especially on glass surfaces. Remove the residues with a soft moist cloth.
  • Page 203: Sterilizing The Protective Cover Of The Smart Device Holder

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.2 Surfaces 5.2.3 Sterilizing the protective cover of the smart device holder The silicone protective mat of the smart device holder can be cleaned with a brush and wipe disinfected.
  • Page 204: Disinfect Handles And Instrument Holders And Sterilize Silicone Mats

    WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! Dentist element 1. Take the silicone mat off the dentist element.
  • Page 205 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.2 Surfaces Assistant element 1. Lift the silicone mat from the assistant element. 2. Take the instruments out of the instrument holders.
  • Page 206: Disinfecting The Tray

    WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! 1. Take the silicone mat off the tray. 2. Hold the tray tightly.
  • Page 207: Care For, Clean, And Disinfect Upholstery

    5.2 Surfaces 5.2.7 Care for, clean, and disinfect upholstery Special care, cleaning, and disinfecting agents are recommended by Dentsply Sirona for the care, cleaning, and disinfection of the upholstery. WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting...
  • Page 208: Clean The Foot Control

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.2 Surfaces Instructions for Use Treatment Center Axano Pure 5.2.8 Clean the foot control Regular cleaning of the foot control improves its stability. ➢ Clean the anti-slip coating on bottom plate with a moist cloth (water).
  • Page 209: Instruments And Instrument Hoses

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses Instruments and instrument hoses 5.3.1 Purge the water lines (purge function) To reduce bacteria, the water lines of the water-carrying instruments of the dentist element and the Sprayvit multifunctional syringe of the dentist and assistant element are purged with water.
  • Page 210 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure Ä The purge time is displayed in blue on the right side (here 20 sec). It can be changed in the "Settings/Hygiene/Purge &...
  • Page 211 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses Graphic grayed out If the Activate foot pedal graphic is grayed out in the "Purge" dialog on the right side, the treatment center has detected that the waste water cannot drain away.
  • Page 212 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure ● Check whether an endo contra-angle handpiece is attached on the motor. It has no spray and blocks the water flow.
  • Page 213: Automatically Purge Water Lines (Autopurge Function)

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses 5.3.2 Automatically purge water lines (autopurge function) The autopurge function enables automatic purging of all water-carrying instruments in the dentist element, the Sprayvit multifunctional syringe in the dentist and assistant elements, the suction hoses, tumbler filling, cuspidor flushing and external device connection.
  • Page 214 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure Error message: Instruments must be in the correct position to start If this message appears after the autopurge function has been opened, the treatment center has detected that not all of the instruments have been placed in the instrument holders.
  • Page 215 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses When the treatment center is first switched on during the working day, this is interpreted to be autopurge mornings. When it is switched off the same day, this is detected as autopurge evenings.
  • Page 216 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure IMPORTANT Pinching the instrument hoses Be careful not to pinch the instrument hoses when inserting the instruments.
  • Page 217 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses Set the quick couplings of the turbines to maximum water flow. Purging water lines The removed instruments are purged with water for the duration of the set autopurge time.
  • Page 218 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure suction hose cleaning, and cuspidor flushing. On the next day, the autopurge function is automatically performed again on all of the...
  • Page 219: Lubricate, Disinfect/Sterilize The Treatment Instruments

    NOTICE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! All disinfectants must be approved in your country and have proven bactericidal, fungicidal and virucidal properties. Use only disinfectants with no protein-fixing properties.
  • Page 220 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure Conduct pre-disinfection ü Wear suitable protective clothing. ü All disinfectants must be approved in your country and have proven bactericidal, fungicidal, and virucidal properties.
  • Page 221 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses 3. Blow the nozzle with 2.5 - 3 bar until no more moisture exits, but at least 10 seconds.
  • Page 222: Disinfecting/Sterilizing Motors And Adapters

    NOTICE Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! All disinfectants must be approved in your country and have proven bactericidal, fungicidal and virucidal properties. Use only disinfectants with no protein-fixing properties.
  • Page 223 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses NOTICE Never clean in an ultrasonic bath! Never immerse in disinfectants! NOTICE Never lubricate the motors! ➢ Remove the handpieces from the motors at the end of the working day so that no oil can run in overnight.
  • Page 224 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure We recommend using Dentsply Sirona DAC Universal for automated cleaning, disinfection, and care. For further details, please refer to the instructions for use supplied with the unit.
  • Page 225 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses 1. When loading the strainer, note the position of the holder to ensure every surface is cleaned.
  • Page 226 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.3 Instruments and instrument hoses Instructions for Use Treatment Center Axano Pure 1. Moisten a clean, lint-free cloth with disinfectant. 2. Wipe the motor/adapter with the moist cloth. When doing so, also wipe any hard-to-reach places.
  • Page 227: Clean And Disinfect/Sterilize The Components Of The Apexlocator

    Clean and disinfect/sterilize the components of the ApexLocator WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! Observe the "General reprocessing instructions"  [→ 196] when reprocessing the ApexLocator components.
  • Page 228: Cleaning/Disinfecting The Sirocam Af / Af+ Intraoral Camera

    Use protective covers. WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! NOTICE The lens window is sensitive to scratches. Deep scratches in the lens window impair image quality.
  • Page 229: Maintaining Treatment Instruments

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses 5.3.4 Maintaining treatment instruments 5.3.4.1 Maintenance of the Sprayvit multifunctional syringe 5.3.4.1.1 Replacing O-rings Replace the O-rings every 3 months.
  • Page 230: Maintaining The Motors

    NOTICE An LED is built into the motors. This LED is designed for the service life of the motor and may be replaced only by Dentsply Sirona or a service technician trained by Dentsply Sirona. Safe operation is no longer guaranteed in case of improper replacement.
  • Page 231: Changing The Cotton Wool Roll On The Turbine Hose

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.3 Instruments and instrument hoses 1. Remove the defective O-rings. 2. Insert the O-rings one after another. Start with the first groove.
  • Page 232: Clean, Disinfect, And Replace Instrument Hoses

    WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! 1. Remove any dirt and disinfectant residues regularly using a mild commercial cleaning agent.
  • Page 233: Suction System

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.4 Suction system Suction system 5.4.1 Cleaning the suction hoses The suction system is exposed daily to secretions, saliva, and blood that contain bacteria. Therefore, for reasons of hygiene, the used suction hoses must be cleaned after each patient, in particular after every treatment involving blood.
  • Page 234 ➢ Do not fill the disinfectant for the water lines into the cleaning agent tank for chemical suction hose cleaning! Use only agents for the suction lines approved by Dentsply Sirona, see "Care, cleaning and disinfecting agents"  [→ 198]. WARNING Wear gloves when performing the following tasks.
  • Page 235 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.4 Suction system 1. Open the maintenance flap on the base of the water unit. The cleaning agent tank for chemical suction hose cleaning is located on the left side.
  • Page 236: Cleaning The Suction System Using The Suction Hose Adapter For The Cuspidor Or Via An External Tank

    WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! NOTICE Domestic cleaning agents foam up. Foaming cleaning agents cause foam and water to be sucked into the dry suction system.
  • Page 237: Cleaning The Suction System Via External Container

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.4 Suction system Cleaning procedure 1. Pour 1 liter of cleaning solution into the cuspidor. 2. Remove the suction hoses from their holders and attach them to the side of the suction hose adapter for the cuspidor as simultaneously as possible.
  • Page 238: Sterilizing, Disinfecting And Lubricating The Suction Handpieces

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.4 Suction system Instructions for Use Treatment Center Axano Pure 5.4.3 Sterilizing, disinfecting and lubricating the suction handpieces Automated reprocessing and sterilization WARNING All parts of the suction handpieces can be automated reprocessed and sterilized.
  • Page 239: Clean The Filter In The Suction Hoses And Disinfect The Suction Hoses

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.4 Suction system 5.4.4 Clean the filter in the suction hoses and disinfect the suction hoses The hoses of the spray aspirator, saliva ejector and surgical suction device as well as the connection hose to the water unit can be detached for rinsing under running water.
  • Page 240 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.4 Suction system Instructions for Use Treatment Center Axano Pure 4. Take the collectors (C) out of the suction hoses. Collect the amalgam residues in a glass filled with water.
  • Page 241: Components Of The Water Unit

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.5 Components of the water unit Components of the water unit 5.5.1 Empty the gold trap The gold trap retains larger solid particles to prevent them from being washed down the drain in the cuspidor bowl.
  • Page 242: Clean/Disinfect The Cuspidor

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Instructions for Use Treatment Center Axano Pure 5.5.2 Clean/disinfect the cuspidor The cuspidor bowl, the tumbler holder (A) and the tumbler outlet (B) can be wiped with surface disinfectants.
  • Page 243: Cleaning The Outlet Lines Of The Cuspidor

    Use the same agent as for the suction lines. WARNING Approved care, cleaning, and disinfecting agents Use only care, cleaning, and disinfecting agents that have been approved by Dentsply Sirona, see "Care, cleaning, and disinfecting agents"  [→ 198]! NOTICE Domestic cleaning agents foam up.
  • Page 244 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Instructions for Use Treatment Center Axano Pure 3. Rinse the cleaning solution away. To do this, pour at least 1 liter of water into the cuspidor.
  • Page 245: Refill Disinfectant For The Water Lines

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.5 Components of the water unit 5.5.4 Refill disinfectant for the water lines The water unit is equipped with a disinfection system. In normal...
  • Page 246 Instructions for Use Treatment Center Axano Pure WARNING Approved care, cleaning, and disinfecting agents Use only the disinfectant for the water lines that has been approved by Dentsply Sirona for the disinfection system; see "Care, cleaning, and disinfecting agents"  [→ 198]. 68 29 514 D3685 D3685.201.02.01.02    2023-12...
  • Page 247: Disinfect Cartridges For Sanitization Adapters And Suction Hose Cleaning

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.5 Components of the water unit 5.5.5 Disinfect cartridges for sanitization adapters and suction hose cleaning The sanitization adapters are placed in a cartridge and can be removed with it from the water unit for cleaning and automated reprocessing.
  • Page 248: Change The Water And Air Filters

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Instructions for Use Treatment Center Axano Pure 5.5.6 Change the water and air filters If you notice any changes in media flows, check that the water and air filters are permeable.
  • Page 249 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.5 Components of the water unit 5. Put the seal (D) back on the filter (E). Then insert both into the screw cap (B ), (C) as shown.
  • Page 250: Changing The Amalgam Rotor

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Instructions for Use Treatment Center Axano Pure 5.5.7 Changing the amalgam rotor Amalgam residues and other solid particles are trapped in the amalgam rotor according to the centrifugal principle.
  • Page 251 6. Place the container in the special packaging and ship it for disposal or authorize a certified waste management company. Installing the amalgam rotor Use only original Dentsply Sirona accessories. Never use a used or recycled amalgam rotor. To reorder the amalgam rotor, see "Spare parts and consumables" ...
  • Page 252 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Instructions for Use Treatment Center Axano Pure 3. Hold the lower part of the amalgam separator (B) so that the latching noses of the bayonet catch are positioned transverse to the water unit.
  • Page 253: Check The Message System Of The Amalgam Separator

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.5 Components of the water unit 5.5.8 Check the message system of the amalgam separator An electronic control device monitors the function of the amalgam separator.
  • Page 254: Emptying The Sediment Container

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Instructions for Use Treatment Center Axano Pure 5.5.9 Emptying the sediment container In addition to other solid particles, a large portion of the amalgam residues are trapped in the sediment container by gravitational force.
  • Page 255 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.5 Components of the water unit 2. Hold the sediment container (B) so that the latching noses of the bayonet catch are positioned transverse to the water unit. Turn the sediment container (B) clockwise to insert it.
  • Page 256: Cleaning The Filter Insert Of The Standard Wet Suction

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.5 Components of the water unit Instructions for Use Treatment Center Axano Pure 5.5.10 Cleaning the filter insert of the standard wet suction With the wet suction system, the automatic separator and amalgam separator or sediment container are not installed in the water unit.
  • Page 257 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.5 Components of the water unit ü The water is completely extracted from the water unit. 1. Open the flap (A) on the base of the water unit.
  • Page 258: Sanitization

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitization Instructions for Use Treatment Center Axano Pure Sanitization Please use only approved agents for disinfecting the water lines. 5.6.1 Sanitizing the treatment center Sanitization can be conducted using the disinfection system. During...
  • Page 259 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.6 Sanitization 2. Tap on the "Sanitization" field. Ä The "Sanitization" dialog appears. If the "Start" key is grayed out in the "Sanitization" dialog, the treatment center has detected that the waste water cannot drain away.
  • Page 260 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitization Instructions for Use Treatment Center Axano Pure ➢ Check the instruments in the instrument positions marked with a red exclamation mark. Ä When all of the instruments have been put in the holders, the sanitization program will continue.
  • Page 261 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.6 Sanitization 3. Assistant element side: Attach the suction hoses to the adapters. When using two large and one small suction hose, the small suction hose must be placed in the middle.
  • Page 262 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitization Instructions for Use Treatment Center Axano Pure If the water flow test fails, the affected instrument can be excluded from the purging process. Affected instruments are marked with a red exclamation mark on the Smart Touch.
  • Page 263 Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.6 Sanitization Error message: Insert instruments in the sanitization adapters If an instrument is accidentally deposited in its holder during the reaction time, the message "Put all rinsable instruments into the colored adapters and press start"...
  • Page 264: Read Out Hygiene Data

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.6 Sanitization Instructions for Use Treatment Center Axano Pure Messages for incomplete sanitization Instruments that have been successfully sanitized are marked with a green check mark, instruments not successfully purged with a red exclamation mark.
  • Page 265: Biofilm Removal By The Service Technician

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.6 Sanitization 5.6.3 Biofilm removal by the service technician If the microbiological test of the water from the treatment center does not correspond with the hygiene requirements in spite of regular sanitization of the water lines and/or regular purge/autopurge cycles, the biofilm must be removed using special chemicals.
  • Page 266: Foot Control And Connection Box

    5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.7 Foot control and connection box Instructions for Use Treatment Center Axano Pure Foot control and connection box 5.7.1 Changing the battery of the wireless foot control The wireless foot control is powered by a battery. An empty battery is detected and reported by the system.
  • Page 267: Changing The Fuse Of The External Device Connection

    Dentsply Sirona 5 Care, cleaning and maintenance by the practice team Instructions for Use Treatment Center Axano Pure 5.7 Foot control and connection box 1. If the adjustable feet A are close to the battery cover B, move them outward slightly.
  • Page 268 5 Care, cleaning and maintenance by the practice team Dentsply Sirona 5.7 Foot control and connection box Instructions for Use Treatment Center Axano Pure ü The IEC outlet socket is not supplying any power. 1. Pull the plug of the connected device out of the IEC outlet socket.
  • Page 269: Maintenance By The Service Engineer

    Dentsply Sirona 6 Maintenance by the service engineer Instructions for Use Treatment Center Axano Pure 6.1 Inspection and maintenance Maintenance by the service engineer Inspection and maintenance In order to ensure the operational safety and reliability of your treatment center and to avoid damage due to wear and tear, annual inspection and maintenance must be performed on your treatment center.
  • Page 270: Maintenance Manual

    6 Maintenance by the service engineer Dentsply Sirona 6.3 Maintenance manual Instructions for Use Treatment Center Axano Pure DANGER It is essential to perform safety checks every 2 years in which electrical faults (e.g. defective insulation) in particular can be detected.
  • Page 271: Spare Parts And Consumables

    Dentsply Sirona 7 Spare parts and consumables Instructions for Use Treatment Center Axano Pure Spare parts and consumables Use only original spare parts and original consumables from Dentsply Sirona. Please order the materials listed below from a specialized dental dealer.
  • Page 272 Instructions for Use Treatment Center Axano Pure O-ring set Sprayvit nozzle 41 76 751 1 x attachment tool and 10 x O-rings 5.5x1.03 Dentsply Sirona T1 Spray (6 x 250 ml cans) 59 01 665 BL E motor BL ISO C/E/S sealing washer, green 63 11 240...
  • Page 273: Disposal

    Dentsply Sirona 8 Disposal Instructions for Use Treatment Center Axano Pure 6.3 Maintenance manual Disposal In accordance with Directive 2012/19/EU and national disposal regulations regarding old electrical and electronic devices, please be advised that such items must be disposed of in a special way within the European Union (EU).
  • Page 274: Battery Disposal

    8 Disposal Dentsply Sirona 8.1 Battery disposal Instructions for Use Treatment Center Axano Pure IMPORTANT Operators of equipment with storage functions for customer and patient data are responsible for deleting all personal data before disposing of the equipment. Battery disposal...
  • Page 275: Overview Of All Function Keys

    Dentsply Sirona 9 Overview of all function keys Instructions for Use Treatment Center Axano Pure 9.1 Fixed Keys Overview of all function keys The following contains a brief description of the function keys on the Smart Touch of the dentist element and the user interface of the assistant element to provide a quick overview of the significance of the symbols on the keys.
  • Page 276: Assistant Element

    9 Overview of all function keys Dentsply Sirona 9.1 Fixed Keys Instructions for Use Treatment Center Axano Pure Bell Freely available relay 240 V, 6 A (connected by service technician) This function can be preset as a button or as a switch in the setup dialog.
  • Page 277: Start Dialog

    Dentsply Sirona 9 Overview of all function keys Instructions for Use Treatment Center Axano Pure 9.2 Start dialog Chair programs 1 and 2 (programmable) Headrest Moves the motorized headrest in/out to adjust the patient's height. Timer function Starts the timer on the assistant element. The timers on the assistant element are preset in the setup dialog, see "SettingsOperation/Timer ...
  • Page 278: Instrument Dialog

    9 Overview of all function keys Dentsply Sirona 9.3 Instrument dialog Instructions for Use Treatment Center Axano Pure Adjusting the chair height Fine adjustment The patient chair moves at reduced speed during for the movements tilt headrest, moving headrest up/down, ErgoMotion, and adjusting chair height.
  • Page 279 Dentsply Sirona 9 Overview of all function keys Instructions for Use Treatment Center Axano Pure 9.3 Instrument dialog Ultrasound intensity Adjusts the intensity of the SiroSonic TL ultrasonic handpiece. The set ultrasound intensity is indicated by a gray filling of the key.
  • Page 280: Endodontics Dialog

    9 Overview of all function keys Dentsply Sirona 9.4 Endodontics dialog Instructions for Use Treatment Center Axano Pure External HF surgical unit Protects the treatment center against interference caused by an external HF surgery unit Endodontics dialog Calibrate bur drive A calibration must be performed each time you change or lubricate the contra-angle handpiece.
  • Page 281: Implantology Dialog

    Dentsply Sirona 9 Overview of all function keys Instructions for Use Treatment Center Axano Pure 9.5 Implantology dialog Torque reverse function When the torque reverse function is switched on, the bur drive automatically switches to counterclockwise rotation when the set torque value is reached.
  • Page 282: Sivision Dialog

    9 Overview of all function keys Dentsply Sirona 9.6 Sivision dialog Instructions for Use Treatment Center Axano Pure By touching the Create system button, a customized implantology system can be set up. By touching the Create instrument button, a customized implant instrument can be set up.
  • Page 283 Dentsply Sirona 9 Overview of all function keys Instructions for Use Treatment Center Axano Pure 9.6 Sivision dialog Tiled layout All open image windows are scaled to a uniform size in the display area and arranged without overlapping. Cascaded layout The opened windows are "cascaded", i.e.
  • Page 284: Media Player

    9 Overview of all function keys Dentsply Sirona 9.6 Sivision dialog Instructions for Use Treatment Center Axano Pure 9.6.2 Media Player Access media player Playback/Pause Stop playback Forward/Back Adjust volume Mute Enlarge/reduce view Show previous/next /page Adjust to page size...
  • Page 285 Dentsply Sirona 9 Overview of all function keys Instructions for Use Treatment Center Axano Pure 9.6 Sivision dialog Stop image transmission on patient monitor Switch between single image and overview Automatic focus before triggering Show focus frame Delete current image...
  • Page 286: Index

    Index Dentsply Sirona Instructions for Use Treatment Center Axano Pure Index Assistant element Numeric Handles, 204 4-way control plate, 63 Instrument positions, 38 Maximum load capacity, 131 Accessory, 26 Overview, 38 Adapter, 236 Sprayvit, 92 Additional devices, 43 Suction handpieces, 134, 238 Additional holder, 36, 38 Use of blasting devices, 19 After blow, 169 Water heater, 168...
  • Page 287 Dentsply Sirona Index Instructions for Use Treatment Center Axano Pure Chemical cuspidor cleaning, 177 Standby switch, 49 Chemical suction hose cleaning, 233 Water heater, 168 Chip blower, 87 Dentsply Sirona Product service, 10 Cleaning agent for suction hose cleaning, 234, 250 Deutscher Verein für Gas und Wasser (DVGW) = German Gas and Water Association, 25...
  • Page 288 Index Dentsply Sirona Instructions for Use Treatment Center Axano Pure Filling quantity of the tumbler, 176 Filter Handles on the dentist and assistant element, 204 Air and water, 248 Hands-free operation, 60 Gold trap, 241 Headrest Suction hoses, 239 Disinfecting, 207 Wet suction device, 256 Double articulating headrest, 31, 69 Fine adjustment, 74...
  • Page 289 Dentsply Sirona Index Instructions for Use Treatment Center Axano Pure Positions on the assistant element, 38 Manufacturer's address, 10 Positions on the dentist element, 36 Maximum load capacity Purging (purge function), 209 Assistant element, 131 Save settings, 85, 168, 181 Dentist element, 79 Side holder, 36, 38 Patient chair, 19 SiroSonic TL, 105...
  • Page 290 Index Dentsply Sirona Instructions for Use Treatment Center Axano Pure Safety switch, 66 SiroSonic TL, 105 Shock positioning, 76 Sivision, 148 Stop movement, 67 Sivision video system, 148 Patient monitor, 148 Sivision View, 21, 228 PC, 21, 155 Sivision View video system, 21, 228 PC control, 155 Sivision View Sidexis 4, 155 ApexLocator distance display, 116 Peristaltic pump, 89...
  • Page 291 Dentsply Sirona Index Instructions for Use Treatment Center Axano Pure Disinfection, 239 Filter, 239 Virtual function keys, 54 Support arm, 141, 206 Voltage and current of the instrument light, 188 Surfaces, 201 Surgical light, 145 Warranty, 269 Surgical suction, 134, 135, 238 Water and air connection, 43 Water as coolant, 87 Taking a photo, 151...
  • Page 292 We reserve the right to make any alterations which may be required due to technical improvements. © SIRONA Dental Systems GmbH Sprache: englisch Printed in Germany D3685.201.02.01.02      2023-12 Ä.-Nr.:  000 000 SIRONA Dental Systems GmbH 68 29 514 D3685 Fabrikstraße 31 Order No   ...