External HF surgical device

External HF surgery units may interfere with the treatment center and
patient monitor. The treatment center should therefore be protected
against interference caused by the external HF surgery unit whenever
the latter is used.
Protecting treatment center from external HF surgery unit
ü The expanded start dialog is displayed on the Smart Touch.
1. Tap on the external HF surgery unit key.
Ä If the function is activated, the key is marked orange. You can
now start using the external HF surgery unit.
2. Tap on the external HF surgery unit key again after using it.
Ä Protection against interference caused by the external HF
surgery unit is turned off.
Electromagnetic disturbances can often be reduced by operating the
external HF surgery unit with a neutral electrode.