4 Operation
4.4 Patient chair
76
4.4.9

Configuring chair programs and shock positioning

Chair programs
The six factory set chair programs:
● Mouth rinsing position S
● Entry/exit position 0
● 1 and 2
● 3 and 4
can be individually reprogrammed for each user profile.
The placement of the chair programs in the Start dialog can be changed
in the setup dialog, see "User interface  [→ 183]".
ü The Start dialog is displayed on the Smart Touch.
1. Move the patient chair to the required treatment position; see
"Moving the chair manually"  [→ 73].
2. Tilt the headrest to the required treatment position; see "Tilting the
headrest".
3. Switch the operating light on or off (will be programmed), see
"Switching the operating light on/off"  [→ 145].
4. Press and hold (>2 s) the desired program key (S, 0, 1, 2 or 3, 4).
Ä A beep sounds. Your settings are now stored under the desired
program key.
Note: Chair programs S, 0, 1, 2 can also be programmed on the
assistant element side.
Shock positioning
When the Shock positioning key is pressed, the patient chair
immediately moves to a position suitable for shock positioning of the
patient.
The factory preset position for shock positioning can be reprogrammed.
1. Move the patient chair to the desired position.
2. Press and hold the Shock positioning key (> 2 s).
NOTICE
Program this key exclusively for shock positioning the patient, never
use as a treatment position.
Instructions for Use Treatment Center Axano Pure
D3685.201.02.01.02    2023-12