4.11.5. How to validate the test mode (simulated flow rate output)

Description
 Checks different outputs (LCD indication, analog output, DO output) upon simulating
flow rate outputs.
With the output at the actuated time as an initial value, the output changes up to the
input value (simulated flow rate target value) in a selected TRACKING TIME, and at
the input value, the output value becomes constant.
During the test mode, "T" blinks on the left end of the 1st line of LCD.
Setting content
TEST MODE
INPUT DATA
TRACKING TIME
Settable range
TEST MODE validation : SETTING (valid), NOT USED (invalid)
INPUT DATA
TRACKING TIME
* For setting TRACKING TIME, 0sec is set to the damping (See 4.9.5).
 By performing the operation, the output of analog outputs DO1 and DO2 will be changed depending on the setting.
Check beforehand whether each output can be changed or not.
 Be sure to resume "NOT USED" after the end of test.
Otherwise, the input value output status will be held until power is turned off.
 If "START/RESET" is selected as TOTAL MODE, the total value also changes. Select "STOP" to prevent the total value
change.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Set the simulated flow rate target to 100%, and the tracking time to 100 [s].
(example)
*Set the "FLOW RATE (%)" beforehand. Refer to "4.9.4 Setting the LCD indication".
Key operation
Press the
Press the
ENT
Press the
Press the
ENT
Press the
Press the
ENT
Press the
ENT
Enter "100" by the
Press the
ENT
Press the
Press the
ENT
INF-TN3FLR-E
: Enables or disables the test mode.
: Simulated flow rate target (percentage of MV full scale).
: Time required to attain the simulated flow rate target
(above input value).
: ±120%
: 0 to 999 seconds
Description
key for 4 times to display "MAINTENANCE MODE".
key once to display "RAS INFORMATION".
ENT
key for 5 times to display "TEST MODE".
key once to blink the cursor.
ENT
key once, and select "SETTING".
key once to register "SETTING".
ENT
key once to blink the cursor on the 2nd line.
ENT
and the
key.
key once to register.
ENT
――― "INPUT DATA" has been registered. ―――
key once to display "TRACKING TIME".
key once to blink the cursor on the 2nd line.
ENT
Input value
Initial value
AUTION
Flow rate output
TRACKING TIME
Display
MAINTENANCE MODE
RAS INFORMATION
0000000000000000
TEST MODE
NOT USED
TEST MODE
NOT USED
TEST MODE
SETTING
INPUT DATA
INPUT DATA
INPUT DATA
INPUT DATA
 COMPLETE 
INPUT DATA
TRACKING TIME
TRACKING TIME
Time
0 %
+000 %
+100
%
100 %
0 sec
000 sec
- 65 -