How To Check The Status Output - Fuji Electric FLR-3 Instruction Manual

Ultrasonic flowmeter m-flow pw flow transmitter
Table of Contents

Advertisement

4.11.4. How to check the status output

Description
 Check the status output.
Setting content
ON: Close the contact.
OFF: Open the contact.
 This operation sets DO1 and DO2 the same contact action.
 Before operation, check whether DO output testing is permitted.
For actual keying, refer to the typical operation indicated below. Set the protection to OFF beforehand. (See Section 4.4.1.)
Operation
Check the contact action.
(example)
Key operation
Press the
Press the
ENT
Press the
Press the
ENT
Note) Contact output is displayed at this time.
"OFF" is given at right.
Press the
Press the
ENT
* Check the contact output "ON".
Press the
Press the
ENT
* Check the contact output "OFF".
Press the
ESC
* It returns to contact output at the normal measurement status.
Press the
ESC
measurement mode.
- 64 -
Description
key for 4 times to display "MAINTENANCE MODE".
key once to display "RAS INFORMATION".
ENT
key for 4 times to display "STATUS".
key once to blink the cursor.
ENT
key once, and select "ON".
key once to register "ON".
ENT
――― "ON" has been registered. ―――
key once, and select "OFF".
key once to register "OFF".
ENT
――― "OFF" has been registered. ―――
key once to stop the cursor from blinking.
ESC
key once and then press the
ESC
AUTION
key once to enter the
Display
MAINTENANCE MODE
RAS INFORMATION
000000000000000
DO CHECK
DO CHECK
DO CHECK
DO CHECK
 COMPLETE 
DO CHECK
DO CHECK
DO CHECK
 COMPLETE 
DO CHECK
DO CHECK
0.000
%
0.000
m3
INF-TN3FLR-E
OFF
OFF
ON
ON
OFF
OFF
OFF