4.9.6. Setting the low flow rate cutting

Description
 The output can be cut when the flow rate is too small.
 Effective for indication, analog output (4-20mA) and total operation.
Settable range: 0 to 5 [m/s] in terms of flow velocity.
(Factory set: 0.150 [m
Note 1) As required, set the low flow rate cut because the flow meter may read a
flow rate when the fluid in the piping is moving on account of convection,
etc. even if the valves are closed.
Note 2) The flow rate unit is as selected by "FLOW UNIT" in "MEASURE SETUP".
(See 4.8.2.)
For actual keying, refer to the typical operation indicated below. Set the protection to
OFF beforehand. (See Section 4.4.1.)
Operation
Set the low flow rate cut point to 0.5 [m
(example)
Key operation
Press the
Press the
ENT
Press the
Press the
ENT
Set "0.5" by the
Press the
ENT
Press the
ESC
the measurement mode.
INF-TN3FLR-E
3
/h])
3
/h].
Description
key twice to display "OUTPUT SETUP".
key once to display "ZERO ADJUSTMENT".
ENT
key twice to display "CUT OFF".
key once to blink the cursor.
ENT
key and the
key.
key once to register.
ENT
――― CUT OFF has been registered. ―――
key once and then press the
ESC
OUTPUT SETUP
ZERO ADJUSTMENT
CUT OFF
CUT OFF
CUT OFF
CUT OFF
CUT OFF
key for 3 times to enter
Output
Flow rate
Low flow cut setting value
Display
SET ZERO
0.150
m3/h
0000.150
m3/h
0000.150
m3/h
0000.550
m3/h
0000.550
m3/h
0000.500
m3/h
 COMPLETE 
0.500
m3/h
0.000
%
0.000
m3