Checking The Led Lit In Red - Fuji Electric FLR-3 Instruction Manual

Ultrasonic flowmeter m-flow pw flow transmitter
Table of Contents

Advertisement

6.5.1.2. Checking the LED lit in red

Check the LED lit in red, following the procedure shown below.
LED lit in red
Check the RAS information
The 2nd or 3rd
from the left is 1.
0110000000000000
No
The 4th or 5th from
the left is 1.
0001100000000000
No
The 6th from the
left is 1.
0000010000000000
No
The 7th from the
left is 1.
0000001000000000
No
The 8th from the
left is 1.
0000000100000000
No
Error status: Malfunction
Cause and remedy:
LED failure  See "6.5.1.1.Checking the LCD/LED"
and "6.5.7.Remedying a hardware
fault".
- 82 -
YES
Error status: Device error, Ultrasonic send/receive signals cannot be
Cause and remedy:
Abnormality of measuring circuit
 Turn the power OFF then ON again.
 See "6.5.7. Remedying a hardware fault".
YES
Error status: No waves are received. Waves are being searched.
Cause and remedy:
Transmission error due to scale or rust within piping
 Move the mounting position of the detector.
Break of signal cable  Repair or replace the signal cable.
Detector mounting failure  Remount the detector.
YES
Transmitter circuit failure or detector failure
 See "6.5.7. Remedying a hardware fault"
Check AGC_U, AGC_D.
45% or less
No
Check P/H_U, P/H_D.
YES
1
YES
1
The values fluctuate
or is outside the range
of 5528 to 6758.
No
1
collected.
If the instrument does not recover, transmitter circuit is failure.
(Window scanning)
YES
Error status:
Signal error  Weak
configuration error
 Signal position is abnormal.
Signal range over
Cause and remedy:
Detector mounting failure  Remount the detector.
Incorrect detector selection
 Change the setting for the detector selection, and
remount the detector.
Transmission error due to scale or rust within piping
 Move the mounting position of the detector.
Incorrect piping data input
YES
Error status: Received wave is abnormal
 Received wave fluctuates, noise
Cause and remedy:
Bubbles or foreign objects included in the fluid
 Stop the flow, and if the measurement is found to be
normal, there is great possibility that air bubbles are
contained. Remove bubbles in such cases.
Noise
 Check ultrasonic signals for existence of noise using
an oscilloscope.
Remove noise if found by performing grounding, etc.
See "6.5.6.Checking received waveforms".
1
receiving
waveform,
INF-TN3FLR-E
Signal