Solving Problems; Well-Plate Handler Does Not Turn On When Switched On; The Well-Plate Handler Arm Does Not Move Down Far Enough To Depress The Plate Sensor Switch - Agilent Technologies 1260 Infinity User Manual

Sample capacity extension
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

Solving Problems

Well-plate Handler does not turn on when switched on.

Possible Cause
Power cord is not firmly plugged in.
Fuse has blown.
Fuse configuration is not correct.
The Well-plate Handler arm does not move down far enough to
depress the plate sensor switch.
Possible Cause
The vertical offset for the position has been
trained too high.
For all other problems, contact Agilent Technologies Product Support.
Agilent 1260 Infinity Sample Capacity Extension User Manual
Troubleshooting and Error Handling
Solution
Verify power cord is firmly seated in receptacles.
Replace fuse(s).
Refer to
"Changing the Fuse Configuration, if
necessary"
on page 26.
Solution
In the software, adjust and save the vertical
offset for the correct plate type.
Note: Place only one well-plate in a rack or device
position when teaching the vertical offset.
5
87

Advertisement

Table of Contents
loading

Table of Contents