Changing the Fixed Capacitor Setting (Model 041 5 2 MHz
Units Only)
The V arioMatch match network Model 0415 2 MHz unit is equipped with a circuit
board containing additional fixed capacitors You can switch these capacitors in
parallel
to the load capacitor to increase the capacitance.
To Change the Capacitor Settings:
1. Tum off RF from the generator.
2. Disconnect AC MAINS.
3. Disconnect RF in (from the RF-GENERATOR).
4. Disconnect RF
5. Remove the top cover, being careful to not damage the plug leading to the fan:
a. Remove the large top cover (with the fan) from the V arioMatch match network
by unscrewing the twelve screws on the cover. Leave the small top and side
covers in place.
b. Disconnect the plug leading from top lid to the fan.
6. Using an appropriate wrench, open or close the jumper bridge to set the desired
capacitance. This circuit board is located in the RF-section of the VarioMatch
match network on the inner side of the
5-36
out
(PLASMA/LOAD).
cabinet
Installation,
Setup, and Operation
close to CLOAD · The following
57020080-00D