CHANGING THE INTERCONNECTOR
To Change the lnterconnector:
1. Locate the alternate interconnector and bridge. These two pieces were attached to
the unit when shipped from the factory.
lnterconnector
Bridge
Figure
5-15.
2. Disconnect water and drain the unit. See
VarioMatch Match Network" on page
3. Using a 1 2 mm wrench, disconnect the compression nut on the copper tube, so
that the tube is loose.
4. Using the 14 mm wrench, unscrew the nut on the copper tube side.
5. Pull the brass tube through the hole. The plastic tube will stay connected.
57020080-00D
AE VarioMatch
Alternate interconnector and bridge
Installation,
Setup, and Operation
Match Network, VM 1000/1500 Platform