High-Frequency Leakage

LigaSure Leakage
Safety Test in Accordance with IEC601-1
Earth Leakage
Patient Leakage
Sink Current
(Mains to Applied Parts)
Ground Bond Test
Valleylab Exchange
Unit Docked to Valleylab Exchange
USB Port Cover reinstalled
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
LigaSure (measured with
leads recommended by
Covidien)
LigaSure (measured directly
at the system terminals)
Normal Conditions
Open Neutral
Normal Conditions
Open Ground
Normal Conditions
Open Ground
Left Tine
Right Tine
<116 mA
<116 mA
<100 mA
<100 mA
μA
0 to 300
μA
0 to 1000
μA
0 to 10
μA
0 to 50
μA
0 to 10
μA
0 to 50
Accept Y, N, or N/A
Accept Y, N, or N/A
Y/N
High-Frequency Leakage
3-15