Accessories; Output Characteristics; Maximum Output For Ligasure Mode - Covidien Valleylab LS10 Service Manual

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

Accessories

The following accessories are for use with the Valleylab LS10 Generator.
Footswitches
Ligasure single pedal footswitch, LS0300 (purple)
Power
AC Cable 100–120 VAC, type SJT16/3
AC Cable 220–240 VAC, type H05WVVF3G1.0 VD
Equipotential
Equipotential Ground Cable, unshielded
Serial Port
USB A to B Cable

Output Characteristics

Maximum Output for LigaSure Mode

The accuracy of the measured RF power shall be within 20% of the actual real output
power at the rated load or 12 W, whichever is greater for the range of 10 W to rated
power.
Precaution
To avoid injury to the patient or surgical team, use only instruments rated for use at, or greater than,
the maximum peak voltages listed below. For example, bipolar instruments must have voltage
ratings of 250 V peak or greater, as shown in the "Open Circuit Peak Voltage (max)" column.
Mode
LigaSure
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Open
Open
Circuit Peak
Circuit P–P
Voltage
Voltage
(max)
(max)
250 V
500 V
Cord Length
15 ft. or less
15 ft. or less
Cord Length
15 ft. or less
Cord Length
6.6 ft. or less
Rated
Power
Short
Load
(max)
Circuit
(max)
RMS
Current
(max)
30
270 W
5.5 A
Accessories
Duty
Cycle
N/A
2-17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents