Covidien Valleylab LS10 Service Manual page 30

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

Electromagnetic Compatibility (IEC 60601-1-2 and IEC 60601-2-2)
Guidance and manufacturer's declaration - electromagnetic immunity
The generator is intended for use in the electromagnetic environment specified below. The customer
or the user of the system should ensure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: U
2-14
IEC 60601 test
level
+/-6 kV contact
+/-8 kV air
+/-2 kV for
power supply
lines
+/-1 kV for
input/output
lines
+/-1 kV
differential
mode
+/-2 kV common
mode
<5% U
(>95%
T
dip in U
)
T
for 0,5 cycle
40% U
T
(>60% dip in U
for 5 cycles
70% U
T
(>30% dip in U
for 25 cycles
<5% U
T
(>95% dip in U
for 5 sec
3 A/m
is the a.c. mains voltage prior to the application of the test level.
T
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Compliance level
+/-6 kV contact
+/-8 kV air
+/-2 kV for power
supply lines
+/-1 kV for input/
output lines
+/-1 kV differential
mode
+/-2 kV common
mode
<5% U
T
(>95% dip in U
)
T
for 0,5 cycle
40% U
T
)
(>60% dip in U
)
T
T
for 5 cycles
70% U
T
)
(>30% dip in U
)
T
T
for 25 cycles
<5%U
T
)
(>95% dip in U
)
T
T
for 5 sec
3 A/m
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be at
least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the generator requires
continued operation during
power mains interruptions, it
is recommended that the
system be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.