Input Frequency; Input Current; Backup Power; Ecg Blanking - Covidien Valleylab LS10 Service Manual

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

100-120 VAC
Cable - SJT16/3, IEC color code, maximum length 15 ft. (5 m)
Plug - minimum 10 A - 125 VAC
Unit receptacle - IEC female, minimum 10 A - 125 VAC
210-240 VAC
Cable - H05VVF3G1.0 VDE, maximum length 15 ft. (5 m)
Plug - minimum 6 A - 250 VAC
Unit receptacle - IEC female, minimum 6 A - 250 VAC
Important
Contact your local Covidien representative for alternative internationally approved power-cord
options.

Input Frequency

The system operates within specification at all line-input frequencies between 48 Hz and
62 Hz. The user does not need to reconfigure the system for different line frequencies.

Input Current

The generator draws no more than 8 A at input voltages between 100 V and 240 V.

Backup Power

The generator retains all user-programmed features, calibration, and statistical data when
switched off and unplugged. The energy platform operates within specification when
switched over to a supplied-line power by hospital backup systems.

ECG Blanking

An ECG blanking port is provided to signal other devices that the generator is active. The
receptacle is a 2.5 mm mono jack. It is electrically isolated from the internal ground
referenced electronics with the shell electrically connected to the chassis for ESD
protection. The ECG blanking port is rated: 12 VDC @ 0.2A.
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Performance Characteristics
2-7