Optional Service Applications
Sensor +5 V: 5.073
Sensor -5 V: -5.096.
The accuracy of the internal voltages is:
±3% @ 1.8 ± 0.09 VDC rail
±3% @ 3.3 ± 0.17 VDC rail
±4% @ 5.0 ± 0.25 VDC rail
±5% @ 48.0 ± 2.40 VDC rail
±5% @ RF 48.0 (+2.40/-2.60) VDC rail
±4% @ sensor power supply +5 (+0.25/-0.40) VDC rail
±11% @ sensor power supply -5 (+0.40/-0.25) VDC rail

Configuration Information

Information: While this method may be used to retrieve configuration information, it is
recommended to connect the generator to the Valleylab Exchange for complete
configuration information retrieval.
Remove the USB port cover on the rear of the generator.
1.
Connect the AC cable to the generator and power on the generator.
2.
Connect one end of the USB A to B cable to the USB port on the rear panel of the
3.
generator (located behind the removable USB port cover).
Connect the other end of the USB A to B cable to a computer.
4.
Launch the terminal emulation program on the computer. Configure the terminal
5.
emulation program to: baud rate of 115200 bps, 8 data bits, no parity, 1 stop bit and
no flow control.
Hit Enter. This starts Generator Service mode.
6.
From the VLLS10GEN>> prompt, select Configuration from the four options.
7.
At the Configuration>> prompt the generator displays Configuration data.
8.

Retrieving Logs

Information: While this method may be used to retrieve partial logs, it is recommended
to connect the generator to Valleylab Exchange for complete log retrieval.
Remove the USB port cover on the rear of the generator.
1.
Connect the AC cable to the generator and power on the generator.
2.
Connect one end of the USB A to B cable to the USB port on the rear panel of the
3.
generator (located behind the removable USB port cover).
Connect the other end of the USB A to B cable to a computer.
4.
Launch the terminal emulation program on the computer. Configure the terminal
5.
emulation program to: baud rate of 115200 bps, 8 data bits, no parity, 1 stop bit and
no flow control.
3-20