Press MOD.
19.
Verify the MOD decrease LED lights. The pulse amplitude waveform initially decreases in
20.
amplitude and then stabilizes at P-T-P amplitude of approximately half-inch.
a.
b.
c.
d.
Press LIGHT LEVEL.
21.
Verify the decrease LED lights. The pulse amplitude waveform initially flatlines and
22.
then stabilizes at the same P-T-P amplitude of approximately half-inch.
Service Manual
Figure 10-18. LIGHT Test: BPM 60, SpO
Active visual and audio alarms.
Saturation indication between 73 and 77 inclusive.
BPM indication between 57 and 63 inclusive.
Pulse waveform of approximately one-inch P-T-P amplitude. Actual amplitude
may vary but references high pulse amplitude/high light indications.
Active visual and audio alarms.
Saturation indication between 73 and 77 inclusive.
BPM indication between 57 and 63 inclusive.
Pulse waveform of approximately half-inch P-T-P amplitude. Actual amplitude
may vary but references low pulse amplitude/high light patients.
90, MOD High, Light High
2
Performance Tests
10-39