harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation.
Warning
RFID function may be interfered with by other equipment even if that other equipment complies
with CISPR emission requirements (required by clause 5.2.2.5 b in IEC 60601-1-2:2007.
Guidance and manufacturer's declaration - electromagnetic emissions
The Valleylab Generator is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC61000-3-3
The essential performance requirement per IEC 60601-1 does not apply to the generator.
Basic Safety is the performance requirement used during immunity testing.
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Electromagnetic Compatibility (IEC 60601-1-2 and IEC 60601-2-2)
Compliance
Group 1
Class A
Class A
Complies
Electromagnetic
environment - guidance
The Valleylab Single Channel
Vessel Sealing Generator must
emit electromagnetic energy
in order to perform its
intended function. Nearby
electronic equipment may be
affected.
The Valleylab Single Channel
Vessel Sealing Generator is
suitable for use in all
establishments other than
domestic and those directly
connected to the public low-
voltage power supply network
that supplies buildings used
for domestic purposes.