Table 11-10. Earth And Enclosure Leakage Current Specifications - Covidien Nellcor Operator's Manual

Bedside spo2 patient monitoring system
Hide thumbs Also See for Nellcor:
Table of Contents

Advertisement

Condition
Normal
Single Fault
Normal
Single Fault
Condition
Normal
Single Fault
Normal
Single Fault
Operator's Manual
Table 11-10.  Earth and Enclosure Leakage Current Specifications
Earth Leakage Current
AC Line Polarity
Line Cord
Normal
Closed
Open
Closed
Reversed
Closed
Open
Closed
Enclosure Leakage Current
Neutral
AC Line Polarity
Line Cord
Normal
Closed
Open
Closed
Reversed
Closed
Open
Closed
Neutral
Line Cord
IEC 60601-1
Closed
500 µA
Closed
Open
Closed
500 µA
Closed
Open
Power Line
Ground
ANSI/AAMI ES 60601-1: 2005
Closed
Closed
Open
Closed
Closed
Open
Manufacturer's Declaration
IEC 60601-1
ANSI/AAMI
ES 60601-1: 2005
300 µA
1000 µA
300 µA
1000 µA
IEC 60601-1
100 µA
500 µA
100 µA
500 µA
101