Error and Event Code Strings
Device Data Error Invalid L
Data CRC
Device Data Error Invalid
Signature
Device Data Error Invalid
Public Key
Device Data Error Invalid
Device Type
Device Usage Not Allowed By
Tag
Device Usage Not Allowed By
Tracklist
Device is not compatible with
generator
Instrument Inserted
Instrument Removed
Hand Switch Pressed
Hand Switch Released
Foot Switch Pressed
Foot Switch Released
Barcode instrument identified Information only
Foot Switch Single Fault
RF On
Seal Complete
Seal Regrasp
Seal Reactive
Large Phase Step
DSP Primary Restored
DSP Second Restored
DSP Activation Denied
5-14
Invalid RFID data read.
Invalid RFID data read.
Invalid RFID data read.
Incompatible Instrument
inserted.
Incompatible Instrument
inserted.
Incompatible Instrument
inserted.
Incompatible Instrument
inserted.
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Information only
System error.
Information only
Information only
Information only
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Replace with valid instrument. Return for
service if persists.
Replace with valid instrument. Return for
service if persists.
Replace with valid instrument. Return for
service if persists.
Replace Instrument. Return for service if
persists.
Replace Instrument. Return for service if
persists.
Replace Instrument. Return for service if
persists.
Replace Instrument. Return for service if
persists.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Replace footswitch. Return for service if event
persists.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Return for service.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
action needed.