Replacement Procedures
Replacement Procedures
Warning
Electric Shock Hazard To allow stored energy to dissipate after power is disconnected, wait at
least five minutes before replacing parts.
Precaution
The system contains electrostatic-sensitive components. When repairing the system, work at a
static-control workstation. Wear a grounding strap when handling electrostatic-sensitive
components, except when working on an energized system. Handle PCBAs by their non-
conductive edges. Use an antistatic container for transport of electrostatic-sensitive components
and PCBAs.
Notice
Perform all the steps including the recalibration listed below. Failure to recalibrate the system after
replacing components may result in the system becoming inoperable.
To service many of the components of the system, it is necessary to remove the front
panel of the system. The steps listed here are referenced throughout the procedures for
servicing other components.

Replacement Parts

6-2
Component
Front Panel Assy
PCBA Display
PCBA PROG MAIN
Battery - Lithium COIN 3V 16 mm
ASSY/RFID/Barcode Vibe
Power Supply
ASSY RECEPTACLE LIGASURE
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Service Level Required
Routine Maintenance
Routine Maintenance
Calibration, VLEX, Routine
Maintenance, Power
Curve Check
VLEX, Routine Maintenance
VLEX, Routine Maintenance
Calibration, Routine
Maintenance, Power
Curve Check
Routine Maintenance