Instrument Information; Check System Voltages - Covidien Valleylab LS10 Service Manual

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

The sensor calibration is complete. Turn off the power to the generator to save the
21.
calibration parameters. Check the prompt to ensure the calibration was successful. If
unsuccessful, contact Covidien Technical Service.

Instrument Information

Remove the USB port cover on the rear of the generator.
1.
Connect the AC cable to the generator and power on the generator.
2.
Connect one end of the USB A to B cable to the USB port on the rear panel of the
3.
generator (located behind the removable USB port cover).
Connect the other end of the USB A to B cable to a computer.
4.
Launch the terminal emulation program on the computer. Configure the terminal
5.
emulation program to: baud rate of 115200 bps, 8 data bits, no parity, 1 stop bit and
no flow control.
Hit Enter. This starts Generator Service mode.
6.
From the VLLS10GEN>> prompt, select Test Functions from the four options.
7.
From the Test Functions>> prompt, select Instrument Info from the ten options.
8.
At the Instrument Info>> prompt the generator will scan (either barcode or RFID) the
9.
LigaSure instrument inserted into the instrument port and report the instrument's
accessory ID, SKU, and name.

Check System Voltages

Remove the USB port cover on the rear of the generator.
1.
Connect the AC cable to the generator and power on the generator.
2.
Connect one end of the USB A to B cable to the USB port on the rear panel of the
3.
generator (located behind the removable USB port cover).
Connect the other end of the USB A to B cable to a computer.
4.
Launch the terminal emulation program on the computer. Configure the terminal
5.
emulation program to: baud rate of 115200 bps, 8 data bits, no parity, 1 stop bit and
no flow control.
Hit Enter. This starts Generator Service mode.
6.
From the VLLS10GEN>> prompt, select Test Functions from the four options.
7.
From the Test Functions>> prompt, select Voltages from the ten options.
8.
At the Voltages>> prompt the generator will scan internal low voltage power supply
9.
values:
3.3 V: 3.255
1.8 V: 1.812
5 V: 4.930
48 V: 47.567
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Optional Service Applications
3-19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents