Instruction Guide for Cleaning, Testing, and Packaging
Kendall SCD™ 700 Sequential Compression System
Controller and Tubing Set

Advertisement

Table of Contents
loading

Summary of Contents for Covidien Kendall SCD 700 Series

  • Page 1 Instruction Guide for Cleaning, Testing, and Packaging Kendall SCD™ 700 Sequential Compression System Controller and Tubing Set...
  • Page 2: Table Of Contents

    TABLE OF CONTENTS Section 1 – Instructions for UNUSED Units Confirm Kit Materials Apply DME Label to Kendall SCD™ 700 Controller Packaging Instructions Section 2 – Instructions for Customer Returned Units Reference Documents Equipment List Preparation Tubing Set: Visual Inspection Cleaning Leak Test Kendall SCD™...
  • Page 3: Section 1 - Instructions For Unused Units

    Verify that all the following materials are available: Part Description Quantity Part # Vendor Contact Info 1-800-962-9888 New Kendall SCD™ 700 compression system H29525 Covidien www.covidien.com Return Label 1-800-962-9888 Twist Ties, Blue Plastic F150023 Covidien www.covidien.com 1-800-962-9888 Kendall SCD™ 700 controller QuickStart Guide...
  • Page 4: Packaging Instructions

    Section 1 – Instructions for UNUSED Units (continued) PACKAGING INSTRUCTIONS  Re-package the home controller in the polybag and AIR-PAQ™* inflatable packaging. Place the tubing sets into the original package the way it was received (see picture below). AIR-PAQ INFLATABLE PACKAGING TUBING SET TUBING SET...
  • Page 5 Section 1 – Instructions for UNUSED Units (continued) PACKAGING INSTRUCTIONS (continued)  Seal the top of the box and confirm all labels are visible on the outside of the shipping box as shown below: Serial label on the top of the shipping box; Shipping label on the front wrapping on the left side;...
  • Page 6: Section 2 - Instructions For Customer Returned Units

    Kendall SCD™ 700 system Operation and Service Manual EQUIPMENT LIST  Regulated air source (i.e. compressor or factory air)  Tubing set Leak Test Fixture (provided by Covidien)  Known good tubing set PREPARATION  TUBING SET LEAK TEST FIXTURE SETUP: NOTE: This is to be performed upon receipt of the fixture, to confirm acceptable fixture functionality.
  • Page 7 Section 2 – Instructions for Customer Returned Units (continued) PREPARATION (continued)  TUBING SET LEAK TEST (continued):  Turn ball valve to open position, so that the knob position is in-line with valve body (see picture below).  Adjust the input air pressure so that a pressure of between 125mmHg (2.4 psi) and 135mmHg (2.6 psi) is indicated on the fixture’s pressure gauge, and allow to hold for approximately 2 seconds.
  • Page 8: Tubing Set

    Section 2 – Instructions for Customer Returned Units (continued) TUBING SET VISUAL INSPECTION  Inspect the tubing sets for missing, broken or cracked components. If any components are missing, broken or cracked, discard the tubing set.  Inspect the tubing set for condensation in the tubing. If condensation is present in the tubing, discard the tubing set.
  • Page 9 Section 2 – Instructions for Customer Returned Units (continued) TUBING SET (continued) LEAK TEST (continued)  Turn ball valve to open position, so that the knob position is in-line with valve body (see picture below).  Adjust the input air pressure so that a pressure of between 125mmHg (2.4 psi) and 135mmHg (2.6 psi) is indicated on the fixture’s pressure gauge, and allow to hold for approximately 2 seconds.
  • Page 10: Kendall Scd™ 700 Controller

    Inspect the controller for the presence of the serial label on the bottom of the device. It the serial label is not present, or if it is illegible or damaged, send the controller to the Covidien service center for repair (Attachment A).
  • Page 11: Cleaning

    If battery charge level is less than 2/3, connect the controller to AC power, and leave connected to AC power until battery reaches at least 2/3 level.  If operation of the SCD™ 700 controller is not correct as per the above tests, send the controller to the Covidien service center for repair (Attachment A).
  • Page 12: Re-Packaging Procedure

    List of materials required to re-package the kit is as follows: Part Description Quantity Part # Vendor Contact Info 1-800-962-9888 Kendall SCD™ 700 system corrugate case 1029053 Covidien www.covidien.com Kendall SCD™ 700 system tubing sets (for 1-800-962-9888 9528 Covidien replacements) www.covidien.com Kendall SCD™ 700 system hospital DME shipping...
  • Page 13 Section 2 – Instructions for Customer Returned Units (continued) KENDALL SCD™ 700 CONTROLLER (continued) RE-PACKAGING PROCEDURE (continued) PACKAGING INSTRUCTIONS (continued) Verify that the DME supplier contact information label is on the Kendall SCD™ 700 controller. Place the controller into the polybag and use a twist tie to secure. Place the Kendall SCD™...
  • Page 14: Attachment A: Instructions For Returning Damaged Controllers

    Building 4 Service Dock Boulder, Co 80301 COVIDIEN, COVIDIEN with logo, and Covidien logo are U.S. and internationally registered trademarks of Covidien AG. ™* Trademark of its respective owner. Other brands are trademarks of a Covidien company. Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.