Covidien Valleylab LS10 Service Manual page 81

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

Device Data Error Invalid
Device Type
Device Usage Not Allowed By
Tag
Device Usage Not Allowed By
Tracklist
Device is not compatible with
generator
Instrument Inserted
Instrument Removed
Hand Switch Pressed
Hand Switch Released
Foot Switch Pressed
Foot Switch Released
Barcode instrument identified Normal operation.
Foot Switch Single Fault
RF On
Seal Complete
Seal Regrasp
Seal Reactive
Large Phase Step
DSP Primary Restored
DSP Second Restored
DSP Activation Denied
Host Power Down
Host Power Drop Off
Host Software Upgrade
Complete
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Incompatible Instrument
Replace with valid instrument. Return for service
inserted.
if persists.
Incompatible Instrument
Replace with valid instrument. Return for service
inserted.
if persists.
Incompatible Instrument
Replace with valid instrument. Return for service
inserted.
if persists.
Incompatible Instrument
Replace with valid instrument. Return for service
inserted.
if persists.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
needed.
Replace footswitch. Return for service if event
Accessory failure.
persists.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
Event is for information only. No corrective action
Normal operation.
needed.
System error.
Return for service.
Event is for information only. No corrective action
System error.
needed.
Event is for information only. No corrective action
System error.
needed.
Event is for information only. No corrective action
System error.
needed
Event is for information only. No corrective action
System error.
needed
Event is for information only. No corrective action
System error.
needed
Software upgrade
Event is for information only. No corrective action
successful.
needed
Error and Event Code Strings
5-17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents