Performance Characteristics

High Frequency (RF) Leakage Current (IEC 60601-2-2)

LigaSure
leakage

Input Power

120 Volt
Maximum power at nominal line
voltage:
Idle: 35 VA
Seal: 400 VA
Full regulation range:
90 to 130 Vac
Operating Range: 100 to 120 Vac
1) Mains current maximum:
2) Idle: 389 mA
3) Seal: 4.44 A
Mains line frequency range (nominal):
50 to 60 Hz
Fuses (2) – 5 mm x 20 mm 8 A, 250 V
fast blow, high-breaking capacity
Power plug:
3-prong hospital grade connector
6-6
Measured with leads
recommended by Covidien
<116 mA RMS
240 Volt
Maximum power at nominal line
voltage:
Idle: 35 VA
Seal: 400 VA
Full Regulation Range:
180 to 264 Vac
Operating Range: 210 to 240 Vac
1) Mains current maximum:
2) Idle: 194 mA
rms
3) Seal: 2.22 A
rms
Mains line frequency range (nominal):
50 to 60 Hz
Fuses (2) – 5 mm x 20 mm 8 A, 250 V
fast blow, high-breaking capacity
Power plug:
3-prong locally approved connector
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator
Measured directly at the
system terminals
< 100 mA RMS
rms
rms