Advertisement

Host Dosage Power Exceeded
Host Check SPI Ready Fail
hand Switch Stuck
foot switch stuck
Invalid Activation
Invalid Instrument Inserted
Host POST Successful
Host POST Failed
Host Digital POT Calibration
Host RTC OF Bit Set
Device Info UID MSB
Device Info UID LSB
Device Info Schema
Device Info Tag Data Ver
Device Info SKU MSB
Device Info SKU LSB
Device Info Device Type
Device Info UMU And UMPU Information only
Device Country Setting
Device New Timestamp
Device Data Error Invalid UL
Data CRC
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
High power. Hardware failure
Perform Output check. Recalibrate sensors.
or unit needs calibration
Return for service if it persists.
Hardware failure, or DSP
software not executing.
Power cycle. Return for service if it persists.
Release button. Try new instrument. Return for
Handswitch button is stuck.
service if it persists.
Release footswitch button. Try new footswitch.
Footswitch is stuck.
Return for service if it persists.
System error.
Power cycle. Return for service if it persists.
Instrument inserted that is not
compatible with the VLLS10.
Use valid instrument.
POST cycle completed
Event is for information only. No corrective
successfully
action needed.
Power cycle unit. Return for service if an error
Power On Self Test has failed
occurs.
Digital potentiometer auto
Event is for information only. No corrective
calibrated on startup
action needed.
System error.
Power cycle. Return for service if it persists.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
action needed.
Event is for information only. No corrective
Information only
action needed.
Event is for information only. No corrective
Information only
action needed.
Replace with valid instrument. Return for
Invalid RFID data read.
service if persists.