Error and Event Code Strings
Host Software Upgrade Fail
Host Application Secondary
Image Corrupt
Host Boot Primary Image
Corrupt
DSP Software Upgrade
Complete
DSP Software Upgrade Fail
DSP Calibration Complete
Max Phase Check Complete
Max Phase Check Fail
Host current calibration
success
Host current calibration fail
Host VPeak Calibration success
Host VPeak Calibration fail
Host NVStore Primary
configuration data fail
Host Secondary configuration
data fail
Host Config Mgmt Primary NV
read fail
Host EBD transport receiver
overflow
Host EBD session dropped
message
Host EBD session receiver
overflow
5-18
Software upgrade failed.
System error.
System error.
Software upgrade
successful.
Software upgrade failed.
Primary sensor calibration
passed
Maximum phase calibration
passed
Maximum phase calibration
failed
Current calibration passed
Current calibration failed
Voltage peak calibration
passed
Voltage peak calibration
failed
System error.
System error.
System error.
System error.
System error.
System error.
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Retry software upgrade. Return for service if it
persists.
Event is for information only. No corrective action
needed, return for service if it persists.
Event is for information only. No corrective action
needed, return for service if it persists.
Event is for information only. No corrective action
needed
Retry software upgrade. Return for service if it
persists.
Event is for information only. No corrective action
needed
Event is for information only. No corrective action
needed
Return for service if it persists.
Event is for information only. No corrective action
needed
Retry sensor calibration. Return for service if it
persists.
Event is for information only. No corrective action
needed
Retry maximum output calibration. Return for
service if it persists.
Power cycle unit. Return for service if error re
occurs.
Power cycle unit. Return for service if error re
occurs.
Power cycle unit. Return for service if error re
occurs.
Power cycle unit. Return for service if error re
occurs.
Power cycle unit. Return for service if error re
occurs.
Power cycle unit. Return for service if error re
occurs.