Covidien Valleylab LS10 Service Manual page 76

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

Error and Event Code Strings
Host RF State Mismatch
Host Instrument Info Corrupt System error.
Host Bad Fan
Host Receive Invalid IPC evnet
Host Seal Deactivation
Host Country Not Set
Host Region Not Set
Host VIBE Init Failed
Host Rtc Failure
Host Energy Without
Activation
Host Control Core Fail
Host Send Message Fail
Host Receptacle Switch
Failure
Host Thread Init Fail
Host Over voltage
Host critical Data Sonsor Cal
check Fail
Host Critical Data Aduio table
check Fail
Host Configuration Check Fail
Host Config Mgmt Primary
and Backup Fail
Host EBD Session Message
Registration Fail
5-12
System error.
Hardware failure.
System error or incompatible
versions of Host and DSP
code.
System error.
Unit not configured correctly.
Need to dock and configure.
Unit not configured correctly.
Need to dock and configure.
Scanner hardware failure.
Hardware failure.
System error, hardware
failure, or unit needs
calibration.
System error.
System error.
Receptacle hardware failure.
System error.
High voltage. Hardware failure
or unit needs calibration
Hardware failure (NV store).
Hardware failure (NV store).
Hardware failure (NV store) or
improper configuration setup. Power cycle. Return for service if it persists.
Hardware failure (NV store).
System error.
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Power cycle. Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Reprogram Host and DSP software.
Return for service if it persists.
Power cycle. Return for service if it persists.
Dock to VLEX and set country code. Return for
service if it persists.
Dock to VLEX and set country code. Return for
service if it persists.
Power cycle. Return for service if it persists.
Dock to VLEX and set time. Change battery.
Return for service if it persists.
Perform Output check. Recalibrate sensors.
Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Return for service if it persists.
Perform Output check. Recalibrate sensors.
Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Return for service if it persists.
Power cycle. Check compatibility of VLEX and
VLLS10

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents