Covidien Valleylab LS10 Service Manual page 80

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

Error and Event Code Strings
Event Text String
hand Switch Stuck
foot switch stuck
Invalid Activation
Invalid Instrument Inserted
Host POST Successful
Host POST Failed
Host Digital POT Calibration
Host RTC OF Bit Set
Device Info UID MSB
Device Info UID LSB
Device Info Schema
Device Info Tag Data Ver
Device Info SKU MSB
Device Info SKU LSB
Device Info Device Type
Device Info UMU And UMPU
Device Country Setting
Device New Timestamp
Device Data Error Invalid UL
Data CRC
Device Data Error Invalid L Data
CRC
Device Data Error Invalid
Signature
Device Data Error Invalid Public
Key
5-16
Reason for entry
Handswitch button is stuck.
Footswitch is stuck.
System error.
Instrument inserted that is
not compatible with the
VLLS10.
POST cycle completed
successfully
POST has failed
Digital potentiometer auto
calibrated on startup
System error.
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Information only
Invalid RFID data read.
Invalid RFID data read.
Invalid RFID data read.
Invalid RFID data read.
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Corrective action
Release button. Try new instrument. Return for
service if it persists.
Release footswitch button. Try new footswitch.
Return for service if it persists.
Power cycle. Return for service if it persists.
Use valid instrument.
Event is for information only. No corrective action
needed.
Power cycle unit. Return for service if error re
occurs.
Event is for information only. No corrective action
needed, return for service if it persists.
Power cycle. Return for service if it persists.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Event is for information only. No corrective action
needed.
Replace with valid instrument. Return for service
if persists.
Replace with valid instrument. Return for service
if persists.
Replace with valid instrument. Return for service
if persists.
Replace with valid instrument. Return for service
if persists.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents