Generator Front Panel - Covidien Valleylab LS10 Service Manual

Single channel vessel sealing generator
Hide thumbs Also See for Valleylab LS10:
Table of Contents

Advertisement

Generator Front Panel

1
Power button - Power on: Push and release. Power off: Push and hold for 3
1.
seconds.
System error indicator - Lights up as shown upon a system error in the generator.
2.
Restart the generator. If the error occurs again, contact local technical service or
Covidien Technical Service.
System status indicator -
3.
• White:
In motion - System is performing a self test.
Constant - Ready for use, insert LigaSure instrument.
Flashing - System in service mode, power cycle system for clinical use.
Purple - Ready for sealing/seal cycle complete.
• Purple in motion - Sealing in process.
• Amber - Seal cycle incomplete alert. Inspect, regrasp, and reactivate sealing and
complete the seal cycle.
Usage limit indicator - When illuminated, indicates the inserted device has already
4.
been used. It has not been recertified by the original manufacturer.
Instrument status or switch stuck indicator -
5.
Red - Instrument error or hand/footswitch stuck. Instrument may be invalid,
damaged, or incompatible. Check that compatible LigaSure instrument is in proper
working condition and no activation switches are depressed.
Instrument port - Connect LigaSure instruments when the system status indicator is
6.
white
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
3
2
4
Generator Front Panel
5
6
LigaSure
1-3