Output Receptacle Sink Current
Set the DVM to AC volts (200 mV) and connect the leakage current test circuit.
1.
Turn on the system and connect the end of the leakage current test circuit to mains
2.
voltage through a 120
Connect the other side of the IEC leakage load to all of the output receptacles.
3.
Determine the leakage current using the conventional 1
4.
Verify the leakage current is less than or equal to 50 μA. If the leakage current is
5.
greater than 50 μA, contact Covidien Technical Service (see page 7-4).
Ground Bond Testing
Connect the system to a ground bond tester.
1.
Test between the equipotential ground lug on the rear of the system and the supplied
2.
Covidien power cord, or directly to the middle ground pin of the inlet receptacle.
Initiate the test according to IEC standards.
3.
Specifications are 0.2
4.
directly to the middle pin of the inlet receptacle. If the specifications are not met
during the ground bond testing, contact Covidien Technical Service (see page 7-4).

Docking to Valleylab Exchange

The Valleylab Exchange (VLEX) client is used for Remote Device Management on
Covidien's generators. The Remote Device Management includes the following
functionality:
– Download the latest version of the Valleylab Exchange Client for generators
– Retrieve the logs (pre-update error logs, pre-update event logs, post-update error
logs, post-update event logs)
– Perform software updates on the generator
– Upload the logs collected along with the update results to the Enterprise server
Equipment required:
• USB A to B Cable
• Valleylab Exchange software
After completing service or preventive maintenance, the system should be docked to
1.
Valleylab Exchange to log any changes to the system in the master directory.
The Valleylab Exchange Remote Software System User's Guide is available online at
2.
www.covidien.com/valleylabexchange and contains instructions for docking to
Valleylab Exchange. Follow the steps outlined in the guide for loading Valleylab
Exchange software and docking the system to the Valleylab Exchange.
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual

¾ W resistor.
using a Covidien-supplied power cord or 0.1
Safety Testing in Accordance with IEC601-1
A for each 1 mV.
μ
connected
3-13