Performance Characteristics

High-Frequency (RF) Leakage Current

LigaSure leakage

Input Power

120 Volt
Maximum power at nominal line
voltage:
Idle: 35 VA
Seal: 400 VA
Full regulation range: 90 to 130 Vac
Operating Range: 100 to 120 Vac
Mains current maximum:
Idle: 389 mA
Seal: 4.44 A
Mains line frequency range (nominal)
50 to 60 Hz
Fuses (2) – 5 mm x 20 mm 8 A, 250 V
fast blow, high breaking capacity
Power plug:
3-prong hospital-grade connector

Power Cord Specification

This system is factory equipped with a 220 VAC hospital-grade NEMA 5-15 power cord.
Should the AC power cord need to be replaced to match another plug configuration, the
replacement plug/cable/receptacle configuration must meet or exceed the following
specifications:
2-6
LigaSure (left
tine)
LigaSure (right
tine)
rms
rms
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Measured with leads
recommended by
Covidien
< 116 mA
< 116 mA
240 Volt
Maximum power at nominal line
voltage:
Idle: 35 VA
Seal: 400 VA
Full regulation range: 180 to 240 Vac
Operating Range: 210 to 240 Vac
Mains current maximum:
Idle: 194 mA
rms
Seal: 2.22 A
rms
Mains line frequency range (nominal)
50 to 60 Hz
Fuses (2) – 5 mm x 20 mm 8 A, 250 V
fast blow, High breaking capacity
Power plug:
3-prong locally-approved connector
Measured directly at
the system terminals
< 100 mA
< 100 mA