Guidance and manufacturer's declaration - electromagnetic immunity
The generator is intended for use in the electromagnetic environment specified below. The
customer or the user of the system should assure that it is used in such an environment.
Immunity
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Valleylab LS10, LS Series Single Channel Vessel Sealing Generator Service Manual
Electromagnetic Compatibility (IEC 60601-1-2 and IEC 60601-2-2)
IEC 60601 test
Compliance
level
level
7 V RMS
3 V RMS
150 kHz to
80 MHz
3 V/m
7 V/m
80 MHz to
2.5 GHz
Continued
Electromagnetic environment -
guidance
Portable and mobile RF communications
equipment should be used no closer to
any part of the generator, including
cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of
the transmitter.
Recommended separation distance
d=1.2√ P
d=1.2√P 80 MHz to 800 MHz
d=2.3√P 800 MHz to 2.5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and d is the recommended
separation distance in meters (m).
Field strengths from fixed RF
a
transmitters
, as determined by an
electromagnetic site survey, should be
less than the compliance level in each
b
frequency range
.
Interference may occur in the vicinity of
equipment marked with the following
symbol: