Storage Settings
Start target
z
z
When "Start cond" is "Measure", "Alarm", "UP limit", or "LOW limit", this setting selects which area
should be the reference of the start condition. Select from Area1 to 4, or Calc1 or 2.
Threshold
z
z
When "Start cond" is "UP limit" or "LOW limit", this setting specifies the threshold to be used as the start
condition reference.
5.3.3

Intermittent

Specifies the sampled measurement results and profile storage interval.
0: Stores all acquired data.
1: Repeatedly stores and discards data with each measurement.
This setting affects the storage function only.
Analog output, judgment output, etc. are updated for each sampling.
5.3.4

Repeat

Specifies behavior when the specified number of data items is stored.
● ON: Returns to the beginning and overwrites the currently stored data.
● OFF: Storage is stopped when the preset number of data items is reached. The storage target
automatically changes to OFF.
● Pause: When the start condition is fulfilled, storage is proceeded automatically up to the preset
number of data items and stops. After that, when the start condition is fulfilled again, it starts storage
again from the point it stopped previously. When it reached at upper limit of the storage data, it returns
to the beginning and overwrites the currently stored data same as when this setting is "ON".
Tips
When the area/calculation hold setting is set to other than "None", the calculated storage value will be
different from the displayed value.
Example: Measured value and storage value operation when the hold setting is "Peak"
Measurement
result
HOLD
5-8
Ramco National
Storage value
(7FFF)
800-280-6933 | nsales@ramcoi.com
Measured value
www.Optex-Ramco.com