5.2 Input/Trigger Settings
5.2.1
IN1/IN2/IN3/IN4
Set the operation of the external input terminals. The available selections vary depending on the terminal.
IN1, IN2, and IN3 are used as bits 1 to 3 of a binary number that is used to switch the bank.
Use the sensor menu to switch the functions of terminals that are not used in bank switching.
("5.9.1 Banks" on page 5-38)
IN1
z
z
● BANK1:
IN1 will be used as the first bit for switching banks.
● Reset:
IN1 will be used as the reset input terminal.
IN2
z
z
● BANK2:
IN2 will be used as the second bit for switching banks.
● Hold:
IN2 will be used as the hold input terminal. The set hold operation will be performed for
the areas and calculations whose hold settings have been set to "Sample," "Peak," or
"Bottom" in the Graph/Calc setting ("Calculation settings" in LS-Navigator).
IN3
z
z
● BANK3:
IN3 will be used as the third bit for switching banks.
● Trigger:
IN3 will be used as the trigger input terminal.
IN4
z
z
● Offset:
IN4 will be used as offset input.
● LaserOFF: IN4 will be used as the laser emission stop input. Input is used to stop laser emission and
measurement.
5.2.2
Reset/Inner hold/Inner trig
Reset
z
z
The reset operation is executed from the sensor or from LS-Navigator.
When you apply the reset input, the following values will be cleared.
● The current measured value
● The measured values up to the current point in time (when the moving average has been set)
● The held value
● Storage data
Also, when a reset input is applied, the output will enter trigger input standby state until the measured
value is obtained.