Triggering - The TRIGger Subsystem
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:TRIGger:CONFiguration?
:TRIGger:CONFiguration?
Returns the hardware trigger configuration.
none
DIS:
Trigger connectors are disabled.
DEF:
The Input Trigger Connector is activated, the incoming trigger response for
each slot ":TRIGger[n][:CHANnel[m]]:INPut" on page 119 determines how
each slot responds to an incoming trigger, all slot events (see ":TRIG-
ger[n][:CHANnel[m]]:OUTPut" on page 121) can trigger the Output Trigger
Connector.
PASS:
The same as DEFault but a trigger at the Input Trigger Connector generates a
trigger at the Output Trigger Connector automatically.
LOOP:
The same as DEFault but a trigger at the Output Trigger Connector generates
a trigger at the Input Trigger Connector automatically.
CUSTOM:
A custom configuration is active using either the command ":TRIGger:CON-
Figuration:EXTended" on page 124 or the Agilent 816x VXIplug&play Instru-
ment Driver. See "The Agilent 816x VXIplug&play Instrument Driver" on
page 155.
trig:conf? → DEF<END>
:TRIGger:CONFiguration:FPEDal
:TRIGger:CONFiguration:FPEDal<wsp>OFF|ON|0|1
Enables or disables the Input Trigger connector to be triggered using a Foot Pedal.
A boolean value:
none
trig:conf? → DEF<END>
:TRIGger:CONFiguration:FPEDal?
:TRIGger:CONFiguration:FPEDal?
Returns whether the Input Trigger connector can be triggered using a Foot Pedal.
none
A boolean value:
trig:conf? → DEF<END>

Extended Trigger Configuration

This section includes information for advanced users about how to
customize your use of the trigger system.
OFF or 0: foot pedal disabled (default)
ON or 1: foot pedal enabled
0: foot pedal disabled
1: foot pedal enabled