Agilent Technologies 8163A Programming Manual page 56

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Instrument Setup and Status
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
command:
syntax:
description:
parameters:
response:
example:
56
:STATus:OPERation:ENABle[:LEVel0]?
:STATus:OPERation:ENABle[:LEVel0]?
Returns the OSESM for the OSESR
none
The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
stat:oper:enab? → +128<END>
:STATus:OPERation[:EVENt]:LEVel1?
:STATus:OPERation[:EVENt]:LEVel1?
Returns the Operational Status Event Summary Register (OSESR) for slots 15 to 17 of the
Agilent 8166A Lightwave Multichannel System.
none
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤ 32767):
Bits
15-4
Not used
3
Slot 17 Summary
2
Slot 16 Summary
1
Slot 15 Summary
0
Not used
stat:oper:level1? → +0<END>
:STATus:OPERation:CONDition:LEVel1?
:STATus:OPERation:CONDition:LEVel1?
Returns the Operational Status Condition Summary Register for slots 15 to 17 of the
Agilent 8166A Lightwave Multichannel System.
none
The sum of the results for slots 15 to 17 (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bits
Mnemonics
Agilent 8166A
15-4
Not used
3
Slot 17 Summary
2
Slot 16 Summary
1
Slot 15 Summary
0
Not used
stat:oper:cond:level1? → +0<END>
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Status Reporting – The STATus Subsystem
Mnemonics
Agilent 8166A
Decimal Value
0
8
4
2
0
Decimal Value
0
8
4
2
0