Agilent Technologies 8163A Programming Manual page 77

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Functions – The SENSe Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
Zeros the electrical offsets for a power meter or return loss module.
none
none
sens1:corr:coll:zero
All power meters and return loss modules
Can only be sent to master channel, slave channel is also zeroed.
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO?
:SENSe[n]:[CHANnel[m]]:CORREction:COLLect:ZERO?
Returns the status of the most recent zero command.
none
0:
any other number:
sens1:corr:coll:zero? → 0<END>
All power meters and return loss modules
Master and slave channels are independent.
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
Zeros the electrical offsets for all installed power meter and return loss modules.
none
none
sens:chan:corr:coll:zero:all
All power meters and return loss modules
Command is independent of channel.
zero succeeded without errors.
remote zeroing failed (the number is the error code returned
from the operation).
Measurement Operations & Settings
77

Advertisement

Table of Contents
loading

This manual is also suitable for:

8164a8166a

Table of Contents