Agilent Technologies 8163A Programming Manual page 122

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
122
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm?
Returns the arming response of a channel to an outgoing trigger.
none
A boolean value:
trig1:outp:rearm?
All Agilent 8163A series power meters
Can only be sent to master channel, slave channel parameters are identical.
:TRIGger:CONFiguration
:TRIGger:CONFiguration<wsp><triggering mode>
Sets the hardware trigger configuration with regard to Output and Input Trigger Connec-
tors.
0 or DISabled:
Trigger connectors are disabled.
1 or DEFault:
The Input Trigger Connector is activated, the incoming trigger re-
sponse for each slot ":TRIGger[n][:CHANnel[m]]:INPut" on page 119
determines how each slot responds to an incoming trigger, all slot
events (see ":TRIGger[n][:CHANnel[m]]:OUTPut" on page 121) can
trigger the Output Trigger Connector.
2 or PASSthrough:
The same as DEFault but a trigger at the Input Trigger Connector gen-
erates a trigger at the Output Trigger Connector automatically.
3 or LOOPback:
The same as DEFault but a trigger at the Output Trigger Connector
generates a trigger at the Input Trigger Connector automatically.
none
trig:conf dis
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Triggering - The TRIGger Subsystem
0: trigger rearming disabled (default)
1: trigger rearming enabled.
0<END>