Status Reporting – The STATus Subsystem
command:
syntax:
description:
parameters:
response:
example:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:STATus:QUEStionable[:EVENt][:LEVel0]?
:STATus:QUEStionable[:EVENt][:LEVel0]?
Returns the Questionable Status Event Summary Register (QSESR).
none
The sum of the results for the QSESR as a 16-bit signed integer value (0 .. +32767)
Bits
Agilent 8163A
15
Not used
14
Not used
13
Not used
12
Not used
11
Not used
10
Not used
9
Not used
8
Not used
7
Not used
6
Not used
5
Not used
4
Not used
3
Not used
2
Slot 2 Summary
1
Slot 1 Summary
0
Not used
stat:ques? → +0<END>
Mnemonics
Agilent 8164A
Not used
Not used
Not used
Slot 14 Summary
Not used
Slot 13 Summary
Not used
Slot 12 Summary
Not used
Slot 11 Summary
Not used
Slot 10 Summary
Not used
Slot 9 Summary
Not used
Slot 8 Summary
Not used
Slot 7 Summary
Not used
Slot 6 Summary
Not used
Slot 5 Summary
Slot 4 Summary
Slot 4 Summary
Slot 3 Summary
Slot 3 Summary
Slot 2 Summary
Slot 2 Summary
Slot 1 Summary
Slot 1 Summary
Slot 0 Summary
Level 1 Summary
Instrument Setup and Status
Decimal Val-
ue
Agilent 8166A
0
16384
8192
4096
2048
1024
512
256
128
64
32
16
8
4
2
1