Agilent Technologies 8163A Programming Manual page 90

Lightwave multimeter lightwave measurement system lightwave multichannel system
Hide thumbs Also See for 8163A:
Table of Contents

Advertisement

Measurement Operations & Settings
command:
syntax:
description:
parameters:
response:
example
affects:
command:
syntax:
description:
NO TE
parameters:
response:
example
affects:
command:
syntax:
description:
NO TE
parameters:
response:
example
affects:
90
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
Start the calibration and save the calibration values for a defined termination reference
measurement. See ":SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]" on
page 91 for information on setting the return loss value of your reference reflector.
none
none
sens1:ret:cal:coll:term
All return loss modules
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]<wsp><value>[dB]
Sets the front panel delta, that is, the loss correction value, for example, due to the front
panel connector. Twice this value is added to the measured Return Loss.
Use [l] to set the front panel delta for an external source or the upper or lower wavelength
laser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value of [l].
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
Sets the front panel delta as a float value in dB
none
sens1:ret:cal:corr:fpd 0.08DB
All return loss modules
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]?
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]?
Returns the front panel delta, that is, the loss correction value, for example, due to the
front panel connector. Twice this value is added to the measured Return Loss.
Use [l] to query the front panel delta for an external source or the upper or lower wave-
length laser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value.
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
Returns the front panel delta as a float value in dB
none
sens1:ret:cal:corr:fpd?
All return loss modules
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System,
& Agilent 8166A Lightwave Multichannel System, Third Edition
Measurement Functions – The SENSe Subsystem
+8.00000000E-002<END>