Triggering - The TRIGger Subsystem
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
parameters:
response:
example:
affects:
dual sensors:
command:
syntax:
description:
NO TE
parameters:
NO TE
response:
example:
affects:
dual sensors:
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement System, & Agilent 8166A Lightwave Multichannel
System, Third Edition
:TRIGger[n][:CHANnel[m]]:OUTPut
:TRIGger[n][:CHANnel[m]]:OUTPut
Specifies when an output trigger is generated and arms the module.
DISabled:
Never.
AVGover:
When averaging time period finishes.
MEASure:
When averaging time period begins.
MODulation:
For every leading edge of a digitally-modulated (TTL) signal
STFinished:
When a sweep step finishes.
SWFinished:
When sweep cycle finishes.
SWSTarted:
When a sweep cycle starts.
none
trig1:outp dis
All tunable laser modules and Agilent 8163A series power meters
Can only be sent to master channel, slave channel is also affected.
:TRIGger[n][:CHANnel[m]]:OUTPut?
:TRIGger[n][:CHANnel[m]]:OUTPut?
Returns the condition that causes an output trigger.
none
DISabled:
Never.
AVGover:
When averaging time period finishes.
MEASure:
When averaging time period begins.
MODulation:
For every leading edge of a digitally-modulated (TTL) signal
STFinished:
When a sweep step finishes.
SWFinished:
When sweep cycle finishes.
SWSTarted:
When a sweep cycle starts.
trig1:outp?
dis<END>
All tunable laser modules and Agilent 8163A series power meters
Can only be sent to master channel, slave channel parameters are identical.
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm<wsp>OFF|ON|0|1
Sets the arming response of a channel to an outgoing trigger.
See Table Table 8, for information on how this command affects the generation of
output triggers using power measurements.
A boolean value:
If you return to Local control, all modules return to the default setting.
none
trig1:outp:rearm 1
All Agilent 8163A series power meters
Can only be sent to master channel, slave channel is also affected.
Measurement Operations & Settings
OFF or 0: trigger rearming disabled
ON or 1: trigger rearming enabled (default)